Z-2900-2016Class II

AQUIOS CL Flow Cytometer, Catalog No. B30166; Intended for use with in vitro diagnostic flow cytometry...

Siemens Healthcare Diagnostics Inc. / initiated 2016-09-06 / Terminated / root cause: software

Beckman Coulter Inc. began a recall of AQUIOS CL Flow Cytometer, Catalog No. B30166; Intended for use with in vitro diagnostic flow cytometry applications using up to four florescent detection... on . FDA classified it as Class II and gives the reason as "Beckman Coulter is recalling the AQUIOS CL System because the use of the reagent carousel can lead to misalignment of the carousel and damage to the prep probe". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-2900-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
AQUIOS CL Flow Cytometer, Catalog No. B30166; Intended for use with in vitro diagnostic flow cytometry applications using up to four florescent detection channels.
Product code
OYE Flow Cytometric Reagents And Accessories.
Reason for recall
Beckman Coulter is recalling the AQUIOS CL System because the use of the reagent carousel can lead to misalignment of the carousel and damage to the prep probe.
Root cause tag
software
FDA root cause
Software Design Change
Action
An Urgent Medical Device Recall letter dated 9/6/16 was sent to customers to inform them that the use of Positions 9 and 10 of the reagent carousel can lead to misalignment of the carousel and damage to the prep probe. The letter informs the customers of the impact of the recall, action to be taken, and the resolution by Beckman Coulter. Customers are instructed to complete and return the enclosed Response Form within 10 days. Customers with any questions regarding this notice, are instructed to contact Customer Support Center: http://www.beckmancoulter.com/customersupport/support; call 1-800-369-0333 in the US and Canada; outside the US and Canada, contact their local Beckman Coulter Representative.
Quantity in commerce
148 units total (12 units in US)
Distribution
US: IL, PA, NY, GA, WA, FL, SC. Affected geographies: Angola, Australia, Belgium, Botswana, Canada, Chile, Columbia, Czech Republic, Denmark, Finland, France, Germany, Ghana, Israel, Italy, Malaysia, Nambia, Netherlands, Panama, Poland, Romania, Slovakia, South Africa, Spain, Sweden, Switzerland, Uganda, United Kingdom and Zambia
Product codes and lots
AY20032 AZ19046 AZ12030 AY51072 AZ19047 AZ15044 AY47071 AZ15038 AY47063 AZ12033 AZ19052 AY51075 AY03004 AZ12027 AY25038

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2900-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2900-2016Class IITerminatedVoluntary: Firm initiated