Z-2793-2016Class II

Syngo Plaza picture archiving and communication system (PAC) (PACS) intended to display, process, read...

Siemens Medical Solutions USA, Inc. / initiated 2016-08-02 / Terminated / root cause: labeling and UDI

Siemens Medical Solutions USA, Inc began a recall of Syngo Plaza picture archiving and communication system (PAC) (PACS) intended to display, process, read, report, communicate, distribute, store and archive... on . FDA classified it as Class II and gives the reason as "Siemens is releasing a Customer Advisory Notice to explain meaning of a suffix o.p. that shows on some DICOM images on area, distance or perimeter values. This suffix stands for on projection and indicates that a shown...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-2793-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
Syngo Plaza picture archiving and communication system (PAC) (PACS) intended to display, process, read, report, communicate, distribute, store and archive digital medical images.
Product code
LLG Kit, Igg, Platelet Associated
Reason for recall
Siemens is releasing a Customer Advisory Notice to explain meaning of a suffix o.p. that shows on some DICOM images on area, distance or perimeter values. This suffix stands for on projection and indicates that a shown measurement value may be different from the real value.
Root cause tag
labeling and UDI
FDA root cause
Error in labeling
Action
Siemens mailed to customers on August 04, 2016 a Customer Advisory Notice that provided interpretations about the meaning of measurement values like area, distance and perimeter shown as "o.p." by the syngo.plaza.
Quantity in commerce
101 Systems
Distribution
Nationwide Distribution
Product codes and lots
Model Numbers: 10863171, 10863172, 10863173, 10592457  Serial Numbers:  100262,100389,100198,100258,100259,100538,100135,100133,100213,100214,100031,100032,100484,100362,100233,100140,100141,100142,100143,100144,100416,100105,100106,100109,100110,100111,100112,100113,100114,100115,100116,100117,100119,100120,100461,100457,100245,100447,100364,100314,100298,100299,100386,200277,100139,100027,100028,100300,100559,100440,100441,100197,100170,100025,100026,100084,100186,100189,100190,100191,100192,100195,100196,100550,100033,100034,100306,100307,100232,100449,100450,100185,100596,100256,100199,100188,100193,100194,100257,100226,100228,100425,100029,100030,100253,100203,100204,100132,100247,100248,100434,100215,100491,100495,100230,100221,100212,100539,100206