Z-0002-2017Class II

Merit Inflation Syringe Kit. Catalog Number KOS-02857

Merit Medical Systems, Inc. / initiated 2016-08-25 / Terminated / root cause: labeling and UDI

Merit Medical Systems, Inc. began a recall of Merit Inflation Syringe Kit. Catalog Number KOS-02857 on . FDA classified it as Class II and gives the reason as "Merit Medical Systems, Inc. announces a voluntary field action for Merit Inflation Syring Kit due to unit labels missing required device labeling information". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-0002-2017
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Merit Medical Systems, Inc.
Product
Merit Inflation Syringe Kit. Catalog Number KOS-02857
Product code
MAV Syringe, Balloon Inflation
Reason for recall
Merit Medical Systems, Inc. announces a voluntary field action for Merit Inflation Syring Kit due to unit labels missing required device labeling information.
Root cause tag
labeling and UDI
FDA root cause
Labeling design
Action
Merit Medical Inc. notified all affected customers on August 25, 2016, via verbal and written notification to return the affected units. The returned units will be re-labeled. For further questions, please call (801) 316-3690.
Quantity in commerce
1,297 total units
Distribution
One consignee located in CA. No Canadian, other foreign or VA/govt/military consignees.
Product codes and lots
Lot Numbers: H959038, H970334, H984351.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0002-2017
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0002-2017Class IITerminatedVoluntary: Firm initiated