Z-0316-2017Class II

Safil QUI + UNDY 5/0 (1) 45CM DSMP13 Suture

Aesculap Inc / initiated 2016-08-01 / Terminated / root cause: manufacturing process

Aesculap Implant Systems LLC began a recall of Safil QUI + UNDY 5/0 (1) 45CM DSMP13 Suture on . FDA classified it as Class II and gives the reason as "There is a potential for the needle to detach from the suture thread during surgery. If the needle becomes detached the defective suture will need to be replaced causing the wound to be sutured a second time". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0316-2017
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Aesculap Inc
Product
Safil QUI + UNDY 5/0 (1) 45CM DSMP13 Suture
Product code
GAM Suture, Absorbable, Synthetic, Polyglycolic Acid
Reason for recall
There is a potential for the needle to detach from the suture thread during surgery. If the needle becomes detached the defective suture will need to be replaced causing the wound to be sutured a second time.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
An Urgent Medical Device Recall Notification Letter was mailed to customers on August 11, 2016 asking them to remove and return the affected product immediately.
Quantity in commerce
180 units
Distribution
Distributed in Arizona
Product codes and lots
Lot # 115441 Product Code C1046619

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0316-2017
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0316-2017Class IITerminatedVoluntary: Firm initiated