LipiFlow thermal Pulsation System, consisting of Console and Activator, Model LFD-1100 The LipiFlow Thermal...
Tearscience, Inc began a recall of LipiFlow thermal Pulsation System, consisting of Console and Activator, Model LFD-1100 The LipiFlow Thermal Pulsation System is intended for use by physicians... on . FDA classified it as Class II and gives the reason as "Complaints related to an error condition occurring during patient treatment where the FipiFlow System automatically stopped treatment after detecting a problem with the Activator". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0093-2017
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- TearScience, Inc.
- Product
- LipiFlow thermal Pulsation System, consisting of Console and Activator, Model LFD-1100 The LipiFlow Thermal Pulsation System is intended for use by physicians as a prescription device for the application of localized heat and pressure therapy in adult patients with chronic cystic conditions of the eyelids, including meibomian gland dysfunction (MGD), also known as evaporative dry eye or lipid deficiency dry eye.
- Product code
- ORZ Eyelid Thermal Pulsation System
- Reason for recall
- Complaints related to an error condition occurring during patient treatment where the FipiFlow System automatically stopped treatment after detecting a problem with the Activator.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- The firm, TearScience, sent a "Medical Device Recall" letter dated September 9, 2016 to its consignees/customers on 9/13/2016. The letter described the product, problem and actions to be taken. The customers were instructed to complete and return the enclosed Acknowledgement and Receipt form by fax to (919) 882-9603, email to: [email protected] or mail to TearScience, ATTN: CUSTOMER SERVICE RCL001, 5151 McCrimmon Parkway, Ste 250, Morrisville, NC 27560; to discontinue use of the product; to remove all a product with lot #201615220020 from your inventory; and call TearScience Customer Service at 919-459-4891 to return the product for above mentioned lot# and request new product for replacement. For any questions regarding this letter, please call (919) 459-4891 Monday through Friday, 8:30AM to 5:30PM, Easter Time.
- Quantity in commerce
- 1,200 Activators (120 boxes of 10 units)
- Distribution
- US Distribution including states of: AK, AR, CA, CO, FL, GA, ID, IL, IN, LA, MD, MA, MN, MO, NV, NM, NY, NC, OH, OK, OR, PA, SC, TN, TX, VA , WA, WI and Puerto Rico.
- Product codes and lots
- Lot 201615220020
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0093-2017 | Class II | Terminated | Voluntary: Firm initiated |