Z-0093-2017Class II

LipiFlow thermal Pulsation System, consisting of Console and Activator, Model LFD-1100 The LipiFlow Thermal...

TearScience, Inc. / initiated 2016-09-13 / Terminated / root cause: design

Tearscience, Inc began a recall of LipiFlow thermal Pulsation System, consisting of Console and Activator, Model LFD-1100 The LipiFlow Thermal Pulsation System is intended for use by physicians... on . FDA classified it as Class II and gives the reason as "Complaints related to an error condition occurring during patient treatment where the FipiFlow System automatically stopped treatment after detecting a problem with the Activator". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0093-2017
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
TearScience, Inc.
Product
LipiFlow thermal Pulsation System, consisting of Console and Activator, Model LFD-1100 The LipiFlow Thermal Pulsation System is intended for use by physicians as a prescription device for the application of localized heat and pressure therapy in adult patients with chronic cystic conditions of the eyelids, including meibomian gland dysfunction (MGD), also known as evaporative dry eye or lipid deficiency dry eye.
Product code
ORZ Eyelid Thermal Pulsation System
Reason for recall
Complaints related to an error condition occurring during patient treatment where the FipiFlow System automatically stopped treatment after detecting a problem with the Activator.
Root cause tag
design
FDA root cause
Device Design
Action
The firm, TearScience, sent a "Medical Device Recall" letter dated September 9, 2016 to its consignees/customers on 9/13/2016. The letter described the product, problem and actions to be taken. The customers were instructed to complete and return the enclosed Acknowledgement and Receipt form by fax to (919) 882-9603, email to: [email protected] or mail to TearScience, ATTN: CUSTOMER SERVICE RCL001, 5151 McCrimmon Parkway, Ste 250, Morrisville, NC 27560; to discontinue use of the product; to remove all a product with lot #201615220020 from your inventory; and call TearScience Customer Service at 919-459-4891 to return the product for above mentioned lot# and request new product for replacement. For any questions regarding this letter, please call (919) 459-4891 Monday through Friday, 8:30AM to 5:30PM, Easter Time.
Quantity in commerce
1,200 Activators (120 boxes of 10 units)
Distribution
US Distribution including states of: AK, AR, CA, CO, FL, GA, ID, IL, IN, LA, MD, MA, MN, MO, NV, NM, NY, NC, OH, OK, OR, PA, SC, TN, TX, VA , WA, WI and Puerto Rico.
Product codes and lots
Lot 201615220020

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0093-2017
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0093-2017Class IITerminatedVoluntary: Firm initiated