Z-0011-2017Class II

Ingenuity Core 128 Model Number 728323

Philips North America LLC / initiated 2016-08-24 / Terminated / root cause: software

Philips Medical Systems began a recall of Ingenuity Core 128 Model Number 728323 Product Usage: Computed Tomography X-ray systems intended to product cross-sectional images of the body by computer... on . FDA classified it as Class II and gives the reason as "The infant performance test did not meet the test specification for the Infant Head Uniformity test with High Resolution. As a result, when the user performs High Resolution head scans, the reconstructed images may...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0011-2017
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Philips North America LLC
Product
Ingenuity Core 128 Model Number 728323 Product Usage: Computed Tomography X-ray systems intended to product cross-sectional images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes.
Product code
JAK System, X-Ray, Tomography, Computed
Reason for recall
The infant performance test did not meet the test specification for the Infant Head Uniformity test with High Resolution. As a result, when the user performs High Resolution head scans, the reconstructed images may exhibit degraded image quality manifested as non-uniformity of the image, CT number shift greater than 5 Hounsfield units, and reduction in gray/white matter differentiation. The issue is most sever in the High Resolution mode, and the factory reference pediatric scan protocols are set to High Resolution default. Adults head scans have a factory reference protocol that defaults to standard resolution which doesn't experience the issue.
Root cause tag
software
FDA root cause
Software design
Action
Philips sent an Urgent Field Safety Notice (FSN) CLE 16-045 letter dated August 26, 2016, via a certified letter to customers. The letter identified the affected product, problem, actions to be taken by customer/user and actions planned by Philips to correct the problem. The letter informed customers a Philips field service representative will install a software update on the affected systems. For questions contact your local Philips representative or local Philips Healthcare office. For North America and Canada, contact the Customer Care Solutions Center (1-800-722-9377, follow the prompts)..
Quantity in commerce
524
Distribution
Worldwide Distribution - US Nationwide in the states of:: AK, AL, AR, AZ, CA, CO, CT, DE, GA, IA, IL, IN, FL, KS, KY, LA, MA, MD, ME, MI, MO, MS, MT, NC, ND, NE, NJ, NH, NM, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, VT, WA, WI, WV, & WY. and the countries of Algeria, Argentina, Australia, Bangladesh, Belgium, Bolivia, Brazil, Cambodia, China, COlumbia, Cuba, Denmark, Dijbouti, Dominican REpublic, Egypt, Estonia, Finland, France, Germany, Gilbraltar, Greece, Hong Kong, Hungary, India, Indonesia, Iran, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Korea, Latvia, Lebanon, Libya, Lithuania, Malaysia, Martinique, Monaco, Mongolia, Mozambique, Myanmar, Netherlands, New Zealand, Norway, Palestine, Panama, Peru, Philippines, Poland, Portugal, Qatar, Reunion, Russian Federation, Senegal, Singapore, Slovakia, South Africa, Spain, Sweden, Seitzerland, Taiwan, Tanzania, Thailand, Tunisia, Turkey, Uniter Arab Emirates, United Kingdom, Uzbekistan, and Vietnam.
Product codes and lots
30016, 30023, 30024, 30151, 32000, 32001, 32008, 32010, 32012, 32013, 32015, 32016, 32018, 32021, 32022, 32024, 32026, 32028, 32032, 32033, 32034, 32036, 32037, 32041, 32042, 32043, 32044, 32046, 32047, 32048, 32049, 32050, 32051, 32052, 32053, 32054, 32055, 32056, 32057, 32058, 32059, 32060, 32061, 32062, 32063, 32064, 32065, 32066, 32067, 32068, 32069, 32070, 32071, 32072, 32073, 32074, 32075, 32076, 32077, 32078, 32079, 32080, 32081, 32082, 32084, 32085, 32086, 32087, 32088, 32089, 32090, 32091, 32092, 32093, 32094, 32095, 32096, 32097, 32098, 32099, 32100, 32101, 32102, 32103, 32104, 32105, 32106, 32108, 32109, 32110, 32111, 32112, 32113, 32114, 32115, 32116, 32117, 32118, 32119, 32120, 32121, 32122, 32123, 32124, 32125, 32126, 32127, 32128, 32129, 32130, 32131, 32133, 32134, 32135, 32136, 32137, 32138, 32139, 32140, 32141, 32142, 32143, 32144, 32145, 32146, 32147, 32148, 32149, 32150, 32151, 32152, 32153, 32154, 32155, 32156, 32157, 32158, 32159, 32160, 32161, 32162, 32163, 32164, 32165, 32166, 32167, 32168, 32169, 32171, 32173, 52023, 52028, 52049, 52073,  320001, 320002, 320003, 320005, 320006, 320008, 320010, 320018, 320019, 320024, 320026, 320029, 320030, 320031, 320033, 320037, 320054, 320058, 320059, 320069, 320073, 320074, 320078, 320079, 320080, 320081, 320082, 320083, 320084, 320088, 320090, 320092, 320096, 320099, 320100, 320101, 320103, 320107, 320111, 320114, 320115, 320119, 320122, 320123, 320125, 320126, 320128, 320129, 320131, 320132, 320133, 320134, 320136, 320138, 320144, 320146, 320147, 320200, 320201, 320202, 320203, 320204, 320206, 320207, 320208, 320209, 320210, 320211, 320212, 320213, 320214, 320215, 320216, 320217, 320218, 320219, 320220, 320221, 320222, 320223, 320224, 320225, 320226, 320227, 320228, 320229, 320230, 320231, 320232, 320233, 320234, 320235, 320236, 320237, 320238, 320239, 320240, 320241, 320242, 320243, 320245, 320246, 320247, 320248, 320250, 320251, 320252, 320253, 320254, 320255, 320256, 320257, 320258, 320259, 320260, 320261, 320262, 320263, 320264, 320265, 320266, 320267, 320268, 320269, 320270, 320271, 320272, 320273, 320274, 320275, 320276, 320278, 320279, 320280, 320281, 320282, 320283, 320284, 320285, 320286, 320288, 320289, 320290, 320291, 320292, 320293, 320295, 320296, 320297, 320298, 320299, 320301, 320302, 320303, 320305, 320307, 320309, 320310, 320311, 320312, 320313, 336011, 336012, 336013, 336015, 336016, 336017, 336018, 336023, 336025, 336027, 336028, 336029, 336034, 336040, 336044, 336046, 336051, 336059, 336060, 336064, 336066, 336067, 336068, 336069, 336070, 336071, 336072, 336073, 336075, 336076, 336077, 336078, 336079, 336080, 336081, 336082, 336083, 336084, 336085, 336086, 336087, 336088, 336089, 336090, 336091, 336092, 336093, 336094, 336095, 336096, 336097, 336098, 336099, 336100, 336101, 336102, 336103, 336104, 336105, 336106, 336107, 336108, 336109, 336110, 336111, 336112, 336113, 336114, 336115, 336116, 336117, 336118, 336119, 336120, 336122, 336123, 336124, 336125, 336126, 336127, 336128, 336129, 336130, 336131, 336132, 336133, 336134, 336135, 336136, 336137, 336138, 336139, 336140, 336141, 336142, 336143, 336144, 336145, 336146, 336147, 336148, 336149, 336150, 336151, 336152, 336153, 336154, 336155, 336156, 336157, 336158, 336159, 336160, 336161, 336162, 336163, 336164, 336165, 336166, 336167, 336168, 336169, 336170, 336171, 336172, 336173, 336174, 336175, 336176, 336177, 336178, 336179, 336180, 336181, 336182, 336183, 336184, 336185, 336186, 336187, 336188, 336189, 336190, 336191, 336192, 336193, 336194, 336195, 336196, 336197, 336198, 336199, 336200, 336201, 336202, 336203, 336204, 336205, 336206, 336207, 336208, 336209, 336210, 336211, 336212, 336213, 336214, 336215, 336216, 336217, 336218, 336219, 336220, 336221, 336222, 336223, 336224, 336225, 336226, 336227, 336228, 336229, 336230, 336231, 336232, 336234, 336235, 336236, 336241, 336243, 336245, 336246, 336247, 336248, 336249, 336252, 336256, 336781, 336782, 336783, 336784, 336785, 336786, 336787, 336788, 336789, 336790, 336791

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0011-2017
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0009-2017Class IITerminatedVoluntary: Firm initiated
Z-0010-2017Class IITerminatedVoluntary: Firm initiated
Z-0011-2017Class IITerminatedVoluntary: Firm initiated
Z-0012-2017Class IITerminatedVoluntary: Firm initiated