RayStation 2.5, 3.0, 3.5, 4.0, 4.5, 4.7, 5.0 and 4.3 (InverseArc 1.0) -- Radiation Therapy Treatment Planning...
RAYSEARCH LABORATORIES AB began a recall of RayStation 2.5, 3.0, 3.5, 4.0, 4.5, 4.7, 5.0 and 4.3 (InverseArc 1.0) -- Radiation Therapy Treatment Planning System RayStation is a software system designed... on . FDA classified it as Class II and gives the reason as "RaySearch became aware of the problem as it was discovered in cooperation with a customer experimenting with the dose calculation on a phantom. The correction concerns two issues found with the dose calculation when...". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0038-2017
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- RaySearch Laboratories AB (publ)
- Product
- RayStation 2.5, 3.0, 3.5, 4.0, 4.5, 4.7, 5.0 and 4.3 (InverseArc 1.0) -- Radiation Therapy Treatment Planning System RayStation is a software system designed for treatment planning and analysis of radiation therapy. The treatment plans provide treatment unit set-up parameters and estimates of dose distributions expected during the proposed treatment, and may be used to administer treatments after review and approval by the intended user. The system functionality can be configured based on user needs.
- Product code
- MUJ System, Planning, Radiation Therapy Treatment
- Reason for recall
- RaySearch became aware of the problem as it was discovered in cooperation with a customer experimenting with the dose calculation on a phantom. The correction concerns two issues found with the dose calculation when using a region of interest (ROI) of type Fixation or Support with material override within the patient outline (External ROI) in RayStation 2.5, 3.0, 3.5, 4.0, 4.5, 4.7, 5.0 and 4.3 (InverseArc 1.0).
- FDA root cause
- Under Investigation by firm
- Action
- RaySearch sent a Field Safety Notice, on September 1, 2016, (RSL-D-61-301, dated 9/5/2016), was sent by e-mail to all affected customers. ACTIONS TO BE TAKEN BY THE USER Do not use ROIs of type Fixation or Support within the patient outline. Always use other ROI types for material override within the External ROI. A small overlap with the External ROI is acceptable if the Fixation or Support ROI is mainly outside the External ROI. Please educate planning staff and all users about this workaround. Inspect your product and identify all installed units with the above software version number, then confirm you have read and understood this notice (contact information below). SOLUTION A correction of these issues will be available in the next version of RayStation, scheduled for market release in December 2016. In the meantime, this field safety notice is distributed to all customers. Until a corrected version has been installed, all affected users must maintain awareness of this field safety notice. TRANSMISSION OF THIS FIELD SAFETY NOTICE This notice needs to be passed on to all those who need to be aware within your organization. Please maintain awareness of this notice as long as this version of RayStation is in use to ensure effectiveness of the workaround.
- Quantity in commerce
- USA: 522 units, OUS 846 units
- Distribution
- Nationwide distribution
- Product codes and lots
- Build Numbers: 2.5.0.144, 2.5.1.89, 3.0.0.251, 3.5.0.16, 3.5.1.6, 4.0.0.14, 4.0.1.4, 4.0.2.9, 4.0.3.42, 4.3.0.14, 4.5.0.19, 4.5.1.14, 4.5.2.7, 4.7.0.15, 4.7.1.10, 4.7.2.5, 4.7.4.4, 4.7.5.4, 5.0.0.37, 5.0.1.11 or 5.0.2.35.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0038-2017 | Class II | Terminated | Voluntary: Firm initiated |