Z-2886-2016Class II

Spacelabs Healthcare Xhibit Telemetry Receiver (XTR), Model 96280. Xhibit Telemetry Receivers (XTR) must be...

GE Healthcare / initiated 2016-08-25 / Terminated / root cause: software

Spacelabs Healthcare Inc began a recall of Spacelabs Healthcare Xhibit Telemetry Receiver (XTR), Model 96280. Xhibit Telemetry Receivers (XTR) must be connected to an Xhibit Central Station for the... on . FDA classified it as Class II and gives the reason as "The firm received multiple reports of Xhibit Telemetry System going offline or locking up. In addition, short-duration asystole alarms may not display an audio and visual alarm as reported from a single Canadian hospital". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-2886-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
GE Healthcare
Product
Spacelabs Healthcare Xhibit Telemetry Receiver (XTR), Model 96280. Xhibit Telemetry Receivers (XTR) must be connected to an Xhibit Central Station for the display of patient vital signs.
Product code
MHX Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)
Reason for recall
The firm received multiple reports of Xhibit Telemetry System going offline or locking up. In addition, short-duration asystole alarms may not display an audio and visual alarm as reported from a single Canadian hospital.
Root cause tag
software
FDA root cause
Software design
Action
Spacelabs sent a letter dated August 25, 2016 to affected US consignees informing them of the going offline or locking up issue when Xhibit Central Station is used as part of an Xhibit Telemetry System. On September 1, 2016, a customer letter (translated as necessary) will be emailed to all international subsidiaries and distributors of record. Spacelabs sent the second Urgent - Medical Device Correction letter, dated September 15, 2016, to affected US consignees informing them that they have received reports of certain short-duration asystole alarms generated by the Xhibit Telemetry Receiver (XTR) where there may be no audible or visual alarm. On September 22, 2016, a second letter (translated as necessary) will be emailed to all international subsidiaries and distributors of record. Spacelabs Healthcare will contact consignees to schedule a convenient time for Spacelabs to update the recalled Xhibit Telemetry System at no cost. Customers with Xhibit Telemetry Receivers will get all Xhibit Central Stations updated regardless if connected to an Xhibit Telemetry Receiver. Consignees are directed to technical support for additional information or technical assistance: Technical Support Spacelabs Healthcare, Inc 35301 SE Center St. Snoqualmie, WA 98065 1-800-522-7025 and select 2 for Technical Support
Quantity in commerce
316 XTR units total (248 units in the US and 68 units outside US)
Distribution
Worldwide distribution. US nationwide including Puerto Rico, Australia, Canada, China, France, Germany, Ireland, Italy, Netherlands, Philippines, Poland, Romania, Switzerland, and United Kingdom.
Product codes and lots
Serial numbers of units distributed in the US   6280-000053, 6280-000056, 6280-000057, 6280-000058, 6280-000059,  6280-000066, 6280-000067, 6280-000068, 6280-000077, 6280-000078,  6280-000079, 6280-000083, 6280-000084, 6280-000085, 6280-000086,  6280-000088, 6280-000090, 6280-000092, 6280-000093, 6280-000095,  6280-000096, 6280-000097, 6280-000098, 6280-000099, 6280-000100,  6280-000101, 6280-000102, 6280-000103, 6280-000104, 6280-000105,  6280-000106, 6280-000107, 6280-000108, 6280-000109, 6280-000110,  6280-000111, 6280-000112, 6280-000113, 6280-000114, 6280-000115,  6280-000116, 6280-000117, 6280-000118, 6280-000120, 6280-000121,  6280-000122, 6280-000123, 6280-000124, 6280-000125, 6280-000126,  6280-000127, 6280-000128, 6280-000129, 6280-000131, 6280-000132,  6280-000137, 6280-000138, 6280-000140, 6280-000141, 6280-000142,  6280-000143, 6280-000146, 6280-000147, 6280-000148, 6280-000149,  6280-000151, 6280-000152, 6280-000153, 6280-000154, 6280-000155,  6280-000156, 6280-000158, 6280-000159, 6280-000160, 6280-000161,  6280-000167, 6280-000168, 6280-000170, 6280-000171, 6280-000172,  6280-000173, 6280-000174, 6280-000176, 6280-000177, 6280-000200,  6280-000228, 6280-000229, 6280-000233, 6280-000234, 6280-000235,  6280-000236, 6280-000238, 6280-000239, 6280-000240, 6280-000241,  6280-000242, 6280-000243, 6280-000245, 6280-000250, 6280-000251,  6280-000252, 6280-000253, 6280-000254, 6280-000255, 6280-000256,  6280-000257, 6280-000260, 6280-000261, 6280-000263, 6280-000264,  6280-000265, 6280-000266, 6280-000267, 6280-000268, 6280-000269,  6280-000271, 6280-000272, 6280-000273, 6280-000274, 6280-000275,  6280-000278, 6280-000279, 6280-000280, 6280-000281, 6280-000283,  6280-000284, 6280-000285, 6280-000286, 6280-000287, 6280-000288,  6280-000289, 6280-000298, 6280-000301, 6280-000302, 6280-000303,  6280-000306, 6280-000307, 6280-000308, 6280-000310, 6280-000311,  6280-000312, 6280-000315, 6280-000316, 6280-000318, 6280-000319,  6280-000321, 6280-000322, 6280-000324, 6280-000326, 6280-000327,  6280-000328, 6280-000329, 6280-000332, 6280-000333, 6280-000334,  6280-000338, 6280-000341, 6280-000342, 6280-000343, 6280-000344,  6280-000345, 6280-000346, 6280-000347, 6280-000348, 6280-000349,  6280-000350, 6280-000351, 6280-000352, 6280-000353, 6280-000354,  6280-000355, 6280-000357, 6280-000358, 6280-000363, 6280-000364,  6280-000365, 6280-000366, 6280-000367, 6280-000368, 6280-000369,  6280-000370, 6280-000375, 6280-000376, 6280-000377, 6280-000378,  6280-000380, 6280-000381, 6280-000382, 6280-000383, 6280-000385,  6280-000386, 6280-000387, 6280-000388, 6280-000389, 6280-000390,  6280-000391, 6280-000392, 6280-000394, 6280-000395, 6280-000396,  6280-000397, 6280-000398, 6280-000399, 6280-000402, 6280-000403,  6280-000404, 6280-000407, 6280-000408, 6280-000409, 6280-000410,  6280-000411, 6280-000412, 6280-000413, 6280-000414, 6280-000415,  6280-000417, 6280-000418, 6280-000419, 6280-000420, 6280-000421,  6280-000423, 6280-000426, 6280-000427, 6280-000429, 6280-000430,  6280-000431, 6280-000432, 6280-000436, 6280-000437, 6280-000438,  6280-000439, 6280-000440, 6280-000455, 6280-000456, 6280-000457,  6280-000458, 6280-000459, 6280-000460, 6280-000461, 6280-000462,  6280-000463, 6280-000464, 6280-000465, 6280-000466, 6280-000467,  6280-000468, 6280-000475, 6280-000476.   Serial numbers of units distributed OUTSIDE the US   6280-000048, 6280-000050, 6280-000052, 6280-000060, 6280-000061,  6280-000062, 6280-000063, 6280-000064, 6280-000065, 6280-000076,  6280-000094, 6280-000178, 6280-000179, 6280-000181, 6280-000182,  6280-000183, 6280-000184, 6280-000185, 6280-000186, 6280-000188,  6280-000190, 6280-000191, 6280-000193, 6280-000194, 6280-000195,  6280-000196, 6280-000197, 6280-000198, 6280-000199, 6280-000202,  6280-000225, 6280-000226, 6280-000231, 6280-000232, 6280-000237,  6280-000246, 6280-000247, 6280-000249, 6280-000262, 6280-000270,  6280-000276, 6280-000282, 6280-000290, 6280-000291, 6280-000292,  6280-000293, 6280-000294, 6280-000295, 6280-000297, 6280-000300,  6280-000305, 6280-000317, 6280-000330, 6280-000331, 6280-000356,  6280-000362, 6280-000371, 6280-000372, 6280-000384, 6280-000400,  6280-000401, 6280-000406, 6280-000416, 6280-000424, 6280-000425,  6280-000433, 6280-000435, and 6280-000473.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2886-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2885-2016Class IITerminatedVoluntary: Firm initiated
Z-2886-2016Class IITerminatedVoluntary: Firm initiated