Z-0073-2017Class II
Fujifilm Synapse PACS, Software versions: 4.0.xxx, 4.1.xxx, version 4.2.xxx, version 4.3.xxx Medical imaging...
FUJIFILM Healthcare Americas Corporation / initiated 2016-09-15 / Terminated / root cause: software
Fujifilm Medical Systems U.S.A., Inc. began a recall of Fujifilm Synapse PACS, Software versions: 4.0.xxx, 4.1.xxx, version 4.2.xxx, version 4.3.xxx Medical imaging and information management system, SYNAPSE allows... on . FDA classified it as Class II and gives the reason as "Image data for a patients image may not be correct". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0073-2017
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- FUJIFILM Healthcare Americas Corporation
- Product
- Fujifilm Synapse PACS, Software versions: 4.0.xxx, 4.1.xxx, version 4.2.xxx, version 4.3.xxx Medical imaging and information management system, SYNAPSE allows the archiving and distribution of image information from all modalities
- Product code
- LLZ System, Image Processing, Radiological
- Reason for recall
- Image data for a patients image may not be correct
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- The firm, FMSU, sent a "Medical Device Correction" letter dated September 15, 2016 to its customers. The letter describes the product, problem and actions to be taken. The firm will install a patch that corrects the issue (Synapse IPSS v1.3.1.0 Patch).Customers with Version 4.0 may upgrade to version 4.4 (will not require the patch) or upgrade to 4.1, 4.2, or 4.3 and install the patch. FMSU will install the patch on the server remotely if your site uses Active Monitoring. Your notification that the patch was installed will be your service ticket. For sites not using Active Monitoring, FMSU will schedule a time to install the patch on your server. Questions or concerns, contact the FUJIFILM Customer Support Center at 1-888-FUJIMED (1-800-385-4633) Option #1. International callers can access the FUJIFILM Customer Support Center by calling 1-203-602-3580.
- Quantity in commerce
- 251 units
- Distribution
- US Nationwide Distribution.
- Product codes and lots
- Software versions: 4.0.xxx, 4.1.xxx, version 4.2.xxx, version 4.3.xxx
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0073-2017 | Class II | Terminated | Voluntary: Firm initiated |