Z-1015-2017Class II

Infant Warmer System (IWS)

GE Healthcare / initiated 2016-08-04 / Terminated

GE Healthcare, LLC began a recall of Infant Warmer System (IWS) on . FDA classified it as Class II and gives the reason as "GE Healthcare has recently become aware of a potential safety issue related to loose screws in the Heater Head of the Infant Warmer System (IWS). Hot screws from the Heater Head of the IWS could fall onto the bed if the...". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-1015-2017
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
GE Healthcare
Product
Infant Warmer System (IWS)
Product code
FMT Warmer, Infant Radiant
Reason for recall
GE Healthcare has recently become aware of a potential safety issue related to loose screws in the Heater Head of the Infant Warmer System (IWS). Hot screws from the Heater Head of the IWS could fall onto the bed if the Heater Head assembly has been improperly serviced. This situation can be clinically hazardous because thermal injury to a patient could result. Two injuries have been reported as a result of this issue.
FDA root cause
Use error
Action
This recall strategy includes the depth of the recall, plan for public warning and effectiveness check. The strategy outlined below addresses these elements for this particular recall. The plan for public warning is to contact all customers with potentially affected product by means of an Urgent Medical Device Correction letter. The letter will instruct customers that they can continue to use the device by following instructions on the Addendum to the Service Manual. The Service Manual Addendum provides instructions for checking and tightening the screws. In the US the letter will be sent by a traceable means [either US Postal Service with receipt acknowledgement (registered mail) or equivalent courier service such as FEDEX]. This letter will be sent to: Nurse Manager, Labor & Delivery/NICU; Manager, Bio-Medical Engineering; and Director of Risk Management.
Quantity in commerce
32,073 (US = 24,361; OUS = 7,712)
Distribution
This product has been distributed domestically and internationally .
Product codes and lots
Medical device listing number: D103053, D103054, D103056, D103057, D183529

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1015-2017
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1015-2017Class IITerminatedVoluntary: Firm initiated