GE Healthcare, LLC began a recall of Infant Warmer System (IWS) on . FDA classified it as Class II and gives the reason as "GE Healthcare has recently become aware of a potential safety issue related to loose screws in the Heater Head of the Infant Warmer System (IWS). Hot screws from the Heater Head of the IWS could fall onto the bed if the...". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-1015-2017
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- GE Healthcare
- Product
- Infant Warmer System (IWS)
- Product code
- FMT Warmer, Infant Radiant
- Reason for recall
- GE Healthcare has recently become aware of a potential safety issue related to loose screws in the Heater Head of the Infant Warmer System (IWS). Hot screws from the Heater Head of the IWS could fall onto the bed if the Heater Head assembly has been improperly serviced. This situation can be clinically hazardous because thermal injury to a patient could result. Two injuries have been reported as a result of this issue.
- FDA root cause
- Use error
- Action
- This recall strategy includes the depth of the recall, plan for public warning and effectiveness check. The strategy outlined below addresses these elements for this particular recall. The plan for public warning is to contact all customers with potentially affected product by means of an Urgent Medical Device Correction letter. The letter will instruct customers that they can continue to use the device by following instructions on the Addendum to the Service Manual. The Service Manual Addendum provides instructions for checking and tightening the screws. In the US the letter will be sent by a traceable means [either US Postal Service with receipt acknowledgement (registered mail) or equivalent courier service such as FEDEX]. This letter will be sent to: Nurse Manager, Labor & Delivery/NICU; Manager, Bio-Medical Engineering; and Director of Risk Management.
- Quantity in commerce
- 32,073 (US = 24,361; OUS = 7,712)
- Distribution
- This product has been distributed domestically and internationally .
- Product codes and lots
- Medical device listing number: D103053, D103054, D103056, D103057, D183529
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1015-2017 | Class II | Terminated | Voluntary: Firm initiated |