Z-0051-2017Class II

MicroStic Plus w/SS Wire 5F and MicroStic Plus w/NiTi Wire & Echo Needle 4F. A sheath used to facilitate...

Argon Medical Devices, Inc / initiated 2016-09-08 / Terminated / root cause: manufacturing process

Argon Medical Devices, Inc began a recall of MicroStic Plus w/SS Wire 5F and MicroStic Plus w/NiTi Wire & Echo Needle 4F. A sheath used to facilitate placing a catheter through the skin into a vein or... on . FDA classified it as Class II and gives the reason as "The supplier initiated a recall because affected lots of the micro-introducer have an outer diameter of the dilator hub that is slightly smaller than is capable of securing the dilator to the sheath". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0051-2017
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Argon Medical Devices, Inc
Product
MicroStic Plus w/SS Wire 5F and MicroStic Plus w/NiTi Wire & Echo Needle 4F. A sheath used to facilitate placing a catheter through the skin into a vein or artery.
Product code
DYB Introducer, Catheter
Reason for recall
The supplier initiated a recall because affected lots of the micro-introducer have an outer diameter of the dilator hub that is slightly smaller than is capable of securing the dilator to the sheath
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
An Urgent - Product Recall Notice was sent to consignees identifying the affected device and the reason for the recall. Customers were asked to return all affected devices. The firm will send replacement devices upon receipt of the returned affected product and once assembly of a cleared lot that does not contain recalled devices has been completed. If there are questions regarding the letter or recall action, customers can contact [email protected].
Quantity in commerce
200 units
Distribution
Worldwide Distribution -- US, to the states of TN and MS; and the country of Canada.
Product codes and lots
11149662, 11151447, 11151617, and 11149822.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0051-2017
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0051-2017Class IITerminatedVoluntary: Firm initiated