510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0325-2017Class II

Empowr PS Knee System Box Cut Guide; Used to make the housing resection for the post of the posterior...

Encore Medical, LP / initiated 2016-09-19 / Terminated / root cause: design

Encore Medical, Lp began a recall of Empowr PS Knee System Box Cut Guide; Used to make the housing resection for the post of the posterior stabilized implant on . FDA classified it as Class II and gives the reason as "It has been discovered that during Empowr PS Total Knee replacement, while chiseling for bone removal in the box area, femoral chipping has occurred in surgery. The current design of the captured box cut guides can...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0325-2017
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Encore Medical, LP
Product
Empowr PS Knee System Box Cut Guide; Used to make the housing resection for the post of the posterior stabilized implant.
Product code
JWH Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer
Reason for recall
It has been discovered that during Empowr PS Total Knee replacement, while chiseling for bone removal in the box area, femoral chipping has occurred in surgery. The current design of the captured box cut guides can create an instance where the surgeon is not fully completing the box cut on the side.
Root cause tag
design
FDA root cause
Under Investigation by firm
Action
The recalling firm sent a recall notification letter to affected consignees on 9/19/16. The field safety notice provided a newly revised Surgical Technique to educate users on proper use of the device.
Quantity in commerce
436 units
Distribution
US
Product codes and lots
202351L01, 202434L01, 203779L01, 202351L02, 202434L02, 202351L03, 202434L03, 203779L03, 202351L04, 202434L04, 202351L05, 202434L05, 202351L06, 202434L06, 203779L06, 202351L07, 202434L07, 203779L07, 202351L08, 202434L08, 203779L08, 202351L09, 202434L09, 202351L10, 202434L10, 203203L01

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0325-2017
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0325-2017Class IITerminatedVoluntary: Firm initiated