Empowr PS Knee System Box Cut Guide; Used to make the housing resection for the post of the posterior...
Encore Medical, Lp began a recall of Empowr PS Knee System Box Cut Guide; Used to make the housing resection for the post of the posterior stabilized implant on . FDA classified it as Class II and gives the reason as "It has been discovered that during Empowr PS Total Knee replacement, while chiseling for bone removal in the box area, femoral chipping has occurred in surgery. The current design of the captured box cut guides can...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0325-2017
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Encore Medical, LP
- Product
- Empowr PS Knee System Box Cut Guide; Used to make the housing resection for the post of the posterior stabilized implant.
- Product code
- JWH Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer
- Reason for recall
- It has been discovered that during Empowr PS Total Knee replacement, while chiseling for bone removal in the box area, femoral chipping has occurred in surgery. The current design of the captured box cut guides can create an instance where the surgeon is not fully completing the box cut on the side.
- Root cause tag
- design
- FDA root cause
- Under Investigation by firm
- Action
- The recalling firm sent a recall notification letter to affected consignees on 9/19/16. The field safety notice provided a newly revised Surgical Technique to educate users on proper use of the device.
- Quantity in commerce
- 436 units
- Distribution
- US
- Product codes and lots
- 202351L01, 202434L01, 203779L01, 202351L02, 202434L02, 202351L03, 202434L03, 203779L03, 202351L04, 202434L04, 202351L05, 202434L05, 202351L06, 202434L06, 203779L06, 202351L07, 202434L07, 203779L07, 202351L08, 202434L08, 203779L08, 202351L09, 202434L09, 202351L10, 202434L10, 203203L01