Z-1137-2017Class II

Celerity ECG Cable Accessory Pack, STERILE, Rx ONLY - distributed under the following labels: (a)...

Boston Scientific Corporation / initiated 2016-09-07 / Terminated / root cause: manufacturing process

Angiodynamics Inc. (Navilyst Medical Inc.) began a recall of Celerity ECG Cable Accessory Pack, STERILE, Rx ONLY - distributed under the following labels: (a) angiodynamics UPN/Product No. H787471431, REF/Catalog No... on . FDA classified it as Class II and gives the reason as "The Celerity ECG Cable Accessory Pack has the potential to contain an ECG Clip Cable that was not properly manufactured. The affected product is not soldered per specification, and may result in the device not properly...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1137-2017
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Boston Scientific Corporation
Product
Celerity ECG Cable Accessory Pack, STERILE, Rx ONLY - distributed under the following labels: (a) angiodynamics UPN/Product No. H787471431, REF/Catalog No. 47-143 -- Legal Manufacturer: AngioDynamics, Inc. 10 Glens Falls Technical Park, Glens Falls, NY 12801; and (b) medCOMP UPN/Product No. MRLCA000, REF/Catalog No. MRLCA000 -- medCOMP, 1499 Delp Drive, Harleysville, PA 19438 --- Description: The Celerity ECG Cable Accessory Pack is comprised of two sterile, single use components: a cable cover and an ECG clip cable (alligator clip). The cable cover is used to sleeve the remote control cable component of the Celerity System so that the user can remotely control the monitor within the sterile field. The stereo jack end of the remote control cable pierces the sterile cable cover and interfaces with the stereo jack connector end of the ECG clip cable. The alligator clip side of the ECG clip cable is clamped to the stylet or guidewire to provide the intravascular cavity ECG reading.
Product code
LJW Tracheotome
Reason for recall
The Celerity ECG Cable Accessory Pack has the potential to contain an ECG Clip Cable that was not properly manufactured. The affected product is not soldered per specification, and may result in the device not properly conducting the ECG signal.
Root cause tag
manufacturing process
FDA root cause
Employee error
Action
On 9/7/2016, AngioDynamics, Inc. sent URGENT VOLUNTARY MEDICAL DEVICE RECALL Notifications (dated 9/7/2016) to 47 consignees via Federal Express. Consignees were instructed to segregate and return all affected devices to AngioDynamics and complete and return the Reply Verification Tracking Form, provided in the recall notification. Please call AngioDynamics Customer Service at 1-800-772-6446 Monday - Friday between 8:00 am and 7:00 Pm EST to obtain a replacement or credit for returned product.
Distribution
AZ, CA, FL, GA, IL, IN, KS, MD, MO, MT, NC, NJ, NV, OH, OR, PA, TN, TX, WI
Product codes and lots
AngioDynamics UPN H787471431 Batch/Lot Numbers: 5015180, MBVR240D, MBVT150R, MBVT210D, MBVT220D, MBVT240D, MBVT250D, MBVT680D, MBVT700D, MBVX090D; medCOMP UPN MRLCA000, Batch/Lot Number MBVT240.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1137-2017
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1137-2017Class IITerminatedVoluntary: Firm initiated