Z-0769-2017Class II

Updated Natus neoBLUE blanket LED Phototherapy Systems. It provides phototherapy underneath the infant and...

Natus Medical Incorporated / initiated 2016-09-16 / Terminated / root cause: design

Natus Medical Incorporated began a recall of Updated Natus neoBLUE blanket LED Phototherapy Systems. It provides phototherapy underneath the infant and can be used in a bassinet, open bed, radiant warmer... on . FDA classified it as Class II and gives the reason as "The firm sent the Technical Bulletin to consignees who received the reliability updated neoBLUE blanket LED Phototherapy Systems between March 2016 to August 2016. The reliability update to prevent the early failure of...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0769-2017
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Natus Medical Incorporated
Product
Updated Natus neoBLUE blanket LED Phototherapy Systems. It provides phototherapy underneath the infant and can be used in a bassinet, open bed, radiant warmer, incubator, or while holding the infant. Affected systems including: 1) Updated light boxes (serial number xxxx004283 or greater) have protection built in to turn off the device if an overheat condition is detected. 2) Updated fiberoptic blankets (Lot Nxxxx16-xx or greater) have an improved fiberoptic cable end to reduce the degradation and increase its useful life. Affected Part Numbers and Description: 006244 neoBLUE blanket LED Phototherapy Systems with Large blanket/pad 006254 neoBLUE blanket LED Light Box 006895 neoBLUE blanket LED Phototherapy Systems with Small blanket/pad 007299 neoBLUE blanket LED Phototherapy Systems with Large blanket/pad EUR 007300 neoBLUE blanket LED Phototherapy Systems with Small blanket/pad EUR 007296 neoBLUE blanket LED Phototherapy Systems with Large blanket/pad UK 007298 neoBLUE blanket LED Phototherapy Systems with Small blanket/pad UK 007301 neoBLUE blanket LED Phototherapy Systems with Large blanket/pad AUS 007302 neoBLUE blanket LED Phototherapy Systems with Small blanket/pad AUS
Product code
LBI Unit, Neonatal Phototherapy
Reason for recall
The firm sent the Technical Bulletin to consignees who received the reliability updated neoBLUE blanket LED Phototherapy Systems between March 2016 to August 2016. The reliability update to prevent the early failure of the fiberoptic pads experienced by some customers on older systems. These failures on previous systems involved discoloration and melting of the fiberoptic bundle at the connector that is inserted in the neoBLUE blanket light box.
Root cause tag
design
FDA root cause
Device Design
Action
Natus sent the Urgent Recall Notification/Field Safety Notice and Customer Reply Form (DOC-021042A), dated September 16, 2016 and the Technical Bulletin (DOC-021016) to US consignees via email. The letters and technical bulletin for foreign consignees was sent via email to the appropriate Natus Distribution Partner (DP) on November 7, 2016. US Customers can email the customer reply form to [email protected] or fax to (847-949-6031). US Customers with questions can call Natus Technical Service at 888-496-2887. International customers should contact their local distributor with any questions.
Quantity in commerce
713 units total (403 units in US and 310 units outside US)
Distribution
Worldwide Distribution - USA (nationwide) Distribution to the states of : AL, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MS, NC, ND, NJ, NM, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WI, WV and Puerto Rico. , and to the countries of : Argentina, Austria, Canada, Chile, China, Denmark, Dominican Republic, Finland, France, Germany, Hong Kong, Hungary, Indonesia, Italy, Latvia, Malaysia, Namibia, Paraguay, Poland, Qatar, Romania, Russia, Saudi Arabia, Singapore, Slovenia, South Africa, Spain, Taiwan, Trinidad and Tobago, Turkey, United Arab Emirates, United Kingdom, Uruguay, Viet Nam, and Virgin Islands (British).
Product codes and lots
*************Serial Numbers of Units distributed in the US **************  004329, 004330, 004349, 004360, 004361, 004362, 004363, 004364, 004365, 004366, 004367, 004369, 004370, 004371, 004417, 004421, 004426, 004436, 004437, 004443, 004448, 004449, 004454, 004456, 004473, 004481, 004488, 004490, 004494, 004495, 004502, 004507, 004517, 004523, 004525, 004526, 004527, 004528, 004530, 004533, 004539, 004541, 004544, 004545, 004546, 004548, 004552, 004553, 004554, 004555, 004556, 004557, 004558, 004567, 004568, 004573, 004578, 004581, 004582, 004584, 004588, 004591, 004602, 004608, 004630, 004632, 004637, 004640, 004647, 004659, 004667, 004668, 004670, 004672, 004673, 004677, 004678, 004679, 004681, 004683, 004684, 004685, 004686, 004693, 004694, 004696, 004700, 004706, 004707, 004708, 004709, 004710, 004711, 004712, 004713, 004714, 004715, 004716, 004717, 004718, 004719, 004720, 004721, 004722, 004723, 004724, 004725, 004726, 004727, 004728, 004729, 004730, 004731, 004732, 004733, 004734, 004735, 004736, 004737, 004738, 004739, 004740, 004741, 004742, 004743, 004745, 004746, 004747, 004749, 004750, 004753, 004754, 004756, 004757, 004760, 004761, 004762, 004763, 004764, 004766, 004767, 004768, 004771, 004772, 004773, 004774, 004775, 004776, 004777, 004778, 004779, 004780, 004781, 004782, 004783, 004784, 004785, 004787, 004788, 004789, 004790, 004791, 004793, 004794, 004795, 004796, 004797, 004798, 004800, 004801, 004802, 004803, 004804, 004805, 004806, 004807, 004808, 004810, 004811, 004812, 004814, 004815, 004816, 004818, 004819, 004820, 004822, 004823, 004824, 004825, 004826, 004827, 004828, 004829, 004830, 004831, 004832, 004833, 004834, 004835, 004836, 004837, 004838, 004839, 004840, 004841, 004842, 004843, 004844, 004845, 004846, 004847, 004848, 004849, 004850, 004851, 004852, 004853, 004854, 004855, 004856, 004857, 004858, 004859, 004860, 004861, 004862, 004863, 004864, 004865, 004866, 004867, 004868, 004869, 004870, 004871, 004872, 004873, 004874, 004876, 004877, 004878, 004879, 004880, 004881, 004882, 004884, 004885, 004886, 004887, 004888, 004890, 004891, 004893, 004894, 004895, 004896, 004897, 004899, 004900, 004901, 004902, 004903, 004904, 004905, 004906, 004907, 004909, 004910, 004997, 004998, 004999, 005007, 005008, 005009, 005010, 005011, 005012, 005013, 005014, 005015, 005016, 005017, 005018, 005019, 005020, 005021, 005022, 005023, 005024, 005025, 005026, 005027, 005028, 005029, 005030, 005031, 005032, 005033, 005034, 005035, 005036, 005037, 005039, 005040, 005041, 005042, 005043, 005044, 005045, 005046, 005047, 005048, 005049, 005050, 005051, 005052, 005053, 005054, 005055, 005056, 005057, 005058, 005059, 005060, 005061, 005062, 005063, 005064, 005065, 005066, 005067, 005068, 005069, 005070, 005071, 005072, 005073, 005074, 005075, 005076, 005077, 005078, 005079, 005080, 005081, 005082, 005083, 005084, 005085, 005086, 005087, 005088, 005089, 005090, 005091, 005092, 005093, 005094, 005095, 005096, 005097, 005098, 005099, 005100, 005101, 005103, 005104, 005105, 005106, 005107, 005108, 005112, 005113, 005115, 005117, 005118, 005119, 005120, 005121, 005122, 005123, 005124, 005125, 005128, 005136, 005137, 005138, 005139, 005140, 005141, 005142, 005143, 005144, 005145, 005146, 005147, 005148, 005149, 005150, 005151, 005152, and 005153.   *************Serial Numbers of Units distributed in the US **************    004283, 004284, 004285, 004286, 004287, 004288, 004290, 004291, 004292, 004293, 004294, 004295, 004296, 004298, 004299, 004302, 004303, 004305, 004306, 004307, 004308, 004309, 004310, 004312, 004313, 004314, 004315, 004316, 004319, 004320, 004321, 004322, 004325, 004326, 004327, 004328, 004332, 004333, 004334, 004335, 004337, 004338, 004339, 004340, 004341, 004343, 004344, 004345, 004347, 004348, 004350, 004351, 004353, 004354, 004355, 004356, 004357, 004359, 004372, 004373, 004374, 004375, 004379, 004380, 004381, 004382, 004383, 004384, 004385, 004386, 004388, 004389, 004390, 004391, 004392, 004393, 004394, 004395, 004396, 004397, 004399, 004400, 004401, 004402, 004403, 004404, 004405, 004406, 004408, 004411, 004412, 004413, 004414, 004415, 004416, 004418, 004419, 004420, 004422, 004423, 004424, 004425, 004427, 004428, 004430, 004431, 004432, 004433, 004439, 004440, 004442, 004445, 004446, 004447, 004450, 004451, 004452, 004453, 004455, 004457, 004458, 004459, 004460, 004462, 004463, 004464, 004465, 004466, 004467, 004468, 004469, 004471, 004472, 004474, 004475, 004476, 004477, 004478, 004479, 004480, 004484, 004485, 004486, 004489, 004491, 004492, 004496, 004497, 004498, 004499, 004500, 004501, 004503, 004504, 004505, 004508, 004510, 004511, 004512, 004513, 004514, 004515, 004516, 004518, 004519, 004520, 004521, 004522, 004524, 004531, 004532, 004536, 004542, 004543, 004549, 004551, 004565, 004569, 004570, 004571, 004572, 004574, 004575, 004576, 004577, 004585, 004587, 004589, 004590, 004592, 004593, 004594, 004596, 004597, 004598, 004599, 004600, 004601, 004604, 004605, 004606, 004609, 004610, 004611, 004612, 004613, 004614, 004615, 004616, 004617, 004618, 004619, 004620, 004621, 004622, 004623, 004624, 004625, 004626, 004628, 004629, 004631, 004633, 004634, 004635, 004638, 004639, 004641, 004642, 004643, 004644, 004645, 004646, 004648, 004649, 004650, 004651, 004652, 004653, 004655, 004656, 004657, 004658, 004660, 004662, 004663, 004664, 004665, 004666, 004669, 004671, 004674, 004675, 004680, 004682, 004688, 004689, 004690, 004692, 004695, 004697, 004698, 004699, 004701, 004702, 004703, 004704, 004705, 004744, 004748, 004751, 004752, 004755, 004758, 004765, 004769, 004770, 004792, 004799, 004809, 004813, 004817, 004821, 004875, 004889, 004898, 004908, 005102, 005111, 005114, 005116, 005126, 005127, 005129, 005130, 005131, 005132, 005133, 005134, 005135, 005154, 005155, 005239, 005240, 005241, 4317, 4323, 4324, 4368, and 4377.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0769-2017
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0769-2017Class IITerminatedVoluntary: Firm initiated