510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0111-2017Class II

Vanguard Distal Femoral Augment 75 X 5 RL/LM

Zimmer GmbH / initiated 2016-08-09 / Terminated / root cause: labeling and UDI

Zimmer Biomet, Inc. began a recall of Vanguard Distal Femoral Augment 75 X 5 RL/LM Product Usage: Painful and disabled knee joint resulting from osteoarthritis, rheumatoid arthritis, traumatic... on . FDA classified it as Class II and gives the reason as "Contents of the package do not match the product labeling. Product is labeled: Vanguard Distal Femoral Augment 75 X 5 RL/LM Carton contains: Vanguard Distal Femoral Augment 75 X 5 LL/RM Risks: Adverse health outcomes...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-0111-2017
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Zimmer GmbH
Product
Vanguard Distal Femoral Augment 75 X 5 RL/LM Product Usage: Painful and disabled knee joint resulting from osteoarthritis, rheumatoid arthritis, traumatic arthritis where one or more compartments are involved. Correction of varus, valgus, or posttraumatic deformity. Correction or revision of unsuccessful osteotomy, arthrodesis, or failure of previous joint replacement procedure. The device is a single use implant intended for implantation with bone cement.
Product code
JWH Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer
Reason for recall
Contents of the package do not match the product labeling. Product is labeled: Vanguard Distal Femoral Augment 75 X 5 RL/LM Carton contains: Vanguard Distal Femoral Augment 75 X 5 LL/RM Risks: Adverse health outcomes that could occur as a result of this event are a delay less than 30 minutes and a different size augment being used to complete the procedure. As the augment will not assemble with femoral component due to its shape, it would be easily identified during surgery.
Root cause tag
labeling and UDI
FDA root cause
Labeling mix-ups
Action
On 8/9/2016, URGENT MEDICAL DEVICE RECALL REMOVAL notifications were sent to the affected consignees via courier. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. If after reviewing this notice you have further questions or concerns please call 1-800-348-9500 ext. 1251 between 8:00 am and 5:00pm EST, Monday through Friday. Calls received outside of operating hours will receive a call back the next business day. Alternatively, your questions may be sent by email to [email protected].
Quantity in commerce
5
Distribution
Worldwide Distribution - US Nationwide in the states of NJ, MI, FL and the countries of Sweden and Canada.
Product codes and lots
Item Number 184108 Lot Number 183410 Exp. 8/27/2025

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0111-2017
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0111-2017Class IITerminatedVoluntary: Firm initiated