Z-0378-2017Class II

LFIT Anatomic V40 Femoral Head, Low Friction Ion Treatment, Sterile, 36 mm, REF 6260-9-236; Modular...

Howmedica Osteonics Corp / initiated 2016-08-29 / Terminated

Stryker Howmedica Osteonics Corp. began a recall of LFIT Anatomic V40 Femoral Head, Low Friction Ion Treatment, Sterile, 36 mm, REF 6260-9-236; Modular components designed to be locked onto a femoral hip stem... on . FDA classified it as Class II and gives the reason as "Stryker received several complaints describing incidence of harm secondary to taper lock failure for specific lots of numerous catalog numbers of LFIT Anatomic CoCr V40 Femoral Heads". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0378-2017
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Howmedica Osteonics Corp
Product
LFIT Anatomic V40 Femoral Head, Low Friction Ion Treatment, Sterile, 36 mm, REF 6260-9-236; Modular components designed to be locked onto a femoral hip stem trunnion during surgery for total hip replacement.
Product code
JDI Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Cemented
Reason for recall
Stryker received several complaints describing incidence of harm secondary to taper lock failure for specific lots of numerous catalog numbers of LFIT Anatomic CoCr V40 Femoral Heads.
FDA root cause
Under Investigation by firm
Action
Stryker notified their Branches/Agencies of this recall by e-mail on August 29, 2016 and they were asked to quarantine the affected devices. A Recall Notification Letter and Product Accountability Form was also sent on August 29, 2016 via UPS (with return receipt) to their Branches/Agencies/Hospital Risk Management and Surgeons. On October 11, 2016, Stryker sent an updated recall notification via UPS with return receipt to their affected customers because additional customers and lot numbers were identified.
Quantity in commerce
42,519 units (total Catalog numbers)
Distribution
US Nationwide and Internationally
Product codes and lots
Catalog #6260-9-236 - Head Diameter 36 mm, Offset +5, including all lots manufactured from 1/102 - 7/1/10; Catalog #6260-9-240 - Head Diameter 40 mm, Offset +4, including all lots manufactured from 1/1/06 - 3/4/11; Catalog #6260-9-244 - Head Diameter 44 mm, Offset +4, including all lots manufactured from 1/1/06 - 3/4/11; Catalog #6260-9-340 - Head Diameter 40 mm, Offset +8, including all lots manufactured from 1/1/06 - 3/4/11; Catalog #6260-9-440 - Head Diameter 40 mm, Offset +12, including all lots manufactured from 1/1/06 - 3/4/11; Catalog #6260-9-344 - Head Diameter 44 mm, Offset +8, including all lots manufactured from 1/1/07 - 3/4/11 and Catalog #6260-9-444 - Head Diameter 44 mm, Offset +12, including all lots manufactured from 1/1/06 - 3/4/11.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0378-2017
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0378-2017Class IITerminatedVoluntary: Firm initiated