BD 23 G x .75 in. BD Vacutainer Push Button Blood Collection Set with 12 in. Tubing Catalog # 367342 Intended...
Becton Dickinson & Company began a recall of BD 23 G x .75 in. BD Vacutainer Push Button Blood Collection Set with 12 in. Tubing Catalog # 367342 Intended for venipuncture to obtain blood specimens from... on . FDA classified it as Class II and gives the reason as "BD has received multiple reports of units associated with the 23 G x .75 in. BD Vacutainer Push Button Blood Collection Set with 12 in. Tubing whose lots are exhibiting separation of front and rear barrels upon...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0745-2017
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Becton Dickinson & Co
- Product
- BD 23 G x .75 in. BD Vacutainer Push Button Blood Collection Set with 12 in. Tubing Catalog # 367342 Intended for venipuncture to obtain blood specimens from patients.
- Product code
- JKA Tubes, Vials, Systems, Serum Separators, Blood Collection
- Reason for recall
- BD has received multiple reports of units associated with the 23 G x .75 in. BD Vacutainer Push Button Blood Collection Set with 12 in. Tubing whose lots are exhibiting separation of front and rear barrels upon activation of the safety feature, which retracts the needle.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- The firm, BD, sent a "URGENT PRODUCT RECALL" letter dated 8/8/2016 to its customer with a return response form via UPS 2nd Day mail on 8/8/2016.The letter described the product, problem and actions to be taken. The customers were instructed to immediately review your inventory for the specific lot numbers, quarantine the product and discontinue the shipment of the affected product; complete and return the Recall Response Form by fax to BD at 1-866-913-0661 or email to [email protected] (even if you do not have any product); return all affected products with completed response form per packing instructions; if you have distributed this product, identify your customers and notify them immediately of this product recall; if you would like BD to conduct the notification to your customers, email your customer list within 3 business days to [email protected]. if you have any questions or require assistance with the return of the recalled product and/or availability of credit product, contact 1-888-628-0732 between 8Am and 5 PM ET Monday through Friday.
- Quantity in commerce
- 1,788,150 units
- Distribution
- Worldwide distribution- US (nationwide) and countries of: Canada, Belgium, Brazil.
- Product codes and lots
- 5355531 5355533 5356681 5356682 6006629 6006630 6006631 6006633 6006634 6006650
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0745-2017 | Class II | Ongoing | Voluntary: Firm initiated |