Z-0088-2017Class II

SDMS (SenTec Digital Monitoring System)

Sentec AG / initiated 2016-09-15 / Terminated

SenTec AG began a recall of SDMS (SenTec Digital Monitoring System) on . FDA classified it as Class II and gives the reason as "Sentec AG found that with one batch of an internal SenTec Monitor component (Docking Station Module) there is a risk that a fastening screw could become loose under certain circumstances. A loose screw within the...". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0088-2017
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Sentec AG
Product
SDMS (SenTec Digital Monitoring System)
Product code
LKD Monitor, Carbon-Dioxide, Cutaneous
Reason for recall
Sentec AG found that with one batch of an internal SenTec Monitor component (Docking Station Module) there is a risk that a fastening screw could become loose under certain circumstances. A loose screw within the monitor can create uncontrolled electrical connections, including short circuits, with the risk of smolder or fire and electrical shock (patient and/or operator).
FDA root cause
Component change control
Action
Distributors were sent via e-mail a SenTec 'Mandatory initiated field correction action" letter dated September 15, 2016 . The letter provided a "Description of the issue, Affected products, Identify and localize affected products, Removal from use, Replacement of affected products at end customer & distributor / sub-distributor level". It also provided "action information for sub-distributors and affiliates, and end customers". For questions contact SenTec Inc. by e-mail ([email protected]) or phone (636-343-0000).
Quantity in commerce
75 (54 US; 21 =OUS)
Distribution
Worldwide Distribution - US Distribution to the states of : LA, MI, MO, OH, TN. and to the countries of : United Kingdom, The Netherlands and Australia.
Product codes and lots
SN 306497 to SN 306571.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0088-2017
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0088-2017Class IITerminatedVoluntary: Firm initiated