Z-0088-2017Class II
SDMS (SenTec Digital Monitoring System)
Sentec AG / initiated 2016-09-15 / Terminated
SenTec AG began a recall of SDMS (SenTec Digital Monitoring System) on . FDA classified it as Class II and gives the reason as "Sentec AG found that with one batch of an internal SenTec Monitor component (Docking Station Module) there is a risk that a fastening screw could become loose under certain circumstances. A loose screw within the...". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0088-2017
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Sentec AG
- Product
- SDMS (SenTec Digital Monitoring System)
- Product code
- LKD Monitor, Carbon-Dioxide, Cutaneous
- Reason for recall
- Sentec AG found that with one batch of an internal SenTec Monitor component (Docking Station Module) there is a risk that a fastening screw could become loose under certain circumstances. A loose screw within the monitor can create uncontrolled electrical connections, including short circuits, with the risk of smolder or fire and electrical shock (patient and/or operator).
- FDA root cause
- Component change control
- Action
- Distributors were sent via e-mail a SenTec 'Mandatory initiated field correction action" letter dated September 15, 2016 . The letter provided a "Description of the issue, Affected products, Identify and localize affected products, Removal from use, Replacement of affected products at end customer & distributor / sub-distributor level". It also provided "action information for sub-distributors and affiliates, and end customers". For questions contact SenTec Inc. by e-mail ([email protected]) or phone (636-343-0000).
- Quantity in commerce
- 75 (54 US; 21 =OUS)
- Distribution
- Worldwide Distribution - US Distribution to the states of : LA, MI, MO, OH, TN. and to the countries of : United Kingdom, The Netherlands and Australia.
- Product codes and lots
- SN 306497 to SN 306571.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0088-2017 | Class II | Terminated | Voluntary: Firm initiated |