Agee-WristJack Fracture Reduction System; Item number: CFD-147. The WristJack Fracture Reduction System is a...
Hand Biomechanics Lab Inc began a recall of Agee-WristJack Fracture Reduction System; Item number: CFD-147. The WristJack Fracture Reduction System is a fracture reduction and external fixation system... on . FDA classified it as Class II and gives the reason as "A number of Agee WristJack Fracture Reduction Systems have been manufactured with a molded beam slightly larger than optimal. This may result in the surgeon experiencing difficulty fine-tuning the length (red) and...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0136-2017
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Hand Biomechanics Lab Inc
- Product
- Agee-WristJack Fracture Reduction System; Item number: CFD-147. The WristJack Fracture Reduction System is a fracture reduction and external fixation system for treatment of distal radius fractures. This is a single use system used for a trauma related surgery The WristJack Fracture Reduction System is a fracture reduction and external fixation system for treatment of distal radius fractures. The non-perishable sterile system is packaged in an inner then outer tray and shipped in a cardboard shelf box. Orthopedic:The WristJack Fracture Reduction System is a fracture reduction and external fixation system for treatment of distal radius fractures.
- Product code
- JEC Component, Traction, Invasive
- Reason for recall
- A number of Agee WristJack Fracture Reduction Systems have been manufactured with a molded beam slightly larger than optimal. This may result in the surgeon experiencing difficulty fine-tuning the length (red) and radial/ulnar alignment (blue) adjustments on the device.
- Root cause tag
- manufacturing process
- FDA root cause
- Under Investigation by firm
- Action
- Hand Biomechanics Lab, Inc. sent a Voluntary Medical Device Recall letter to all affected customers beginning on September 14, 2016 and sent by phone/fax numbers - or e-mail. Letters request return of devices which have not yet been used. Customers with questions were instructed to contact Customer Service at (800) 522-5778. For questions regarding this recall call 916-923-5073.
- Quantity in commerce
- 65 units
- Distribution
- Nationwide Distribution
- Product codes and lots
- Lot number: HBL-115-165, serial numbers 47460959 and 47460980; Lot number: HBL-116-014 - all serial numbers; Lot number : HBL-116-040 - 15 systems shipped prior to recall
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0136-2017 | Class II | Terminated | Voluntary: Firm initiated |