Z-0302-2017Class II
ex-TEM Assay for ROTEM delta Thromboelastometry System. Art. no.: 503-05 US, ex-tem. Used to monitor the...
TEM Systems Inc / initiated 2016-09-13 / Terminated / root cause: manufacturing process
TEM Systems Inc began a recall of ex-TEM Assay for ROTEM delta Thromboelastometry System. Art. no.: 503-05 US, ex-tem. Used to monitor the coagulation process via the extrinsic pathway in... on . FDA classified it as Class II and gives the reason as "Customers complained of low/empty reagent vials". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0302-2017
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- TEM Systems Inc
- Product
- ex-TEM Assay for ROTEM delta Thromboelastometry System. Art. no.: 503-05 US, ex-tem. Used to monitor the coagulation process via the extrinsic pathway in citrated whole blood specimens on the ROTEM delta.
- Product code
- JPA System, Multipurpose For In Vitro Coagulation Studies
- Reason for recall
- Customers complained of low/empty reagent vials.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- Customers were notified of the recall by an Urgent Important Customer Information letter on 9/13/2016. The letter described the device and the problem. Customers were told to stop using the product and to call for replacement product. The letter included a Confirmation of Compliance Letter which was to be completed and returned.
- Quantity in commerce
- 637 boxes x 10 vials
- Distribution
- Nationwide Distribution -- AR, CA, DC, GA, HI, ID, IL, IN, LA, ME, MD, MA, MI, MN, MO, NE, NJ, NM, NY, NC, OH, OR, PA, SC, SD, TN, TX, UT, VA, WA, and WV.
- Product codes and lots
- Lot No. 21929007
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0302-2017 | Class II | Terminated | Voluntary: Firm initiated |