Z-0707-2017Class II

Merge, eFilm Workstation and eFilm Lite eFilm Workstation with Modules is a software application that is used...

Merge Healthcare Incorporated / initiated 2016-10-20 / Terminated / root cause: software

Merge Healthcare, Inc. began a recall of Merge, eFilm Workstation and eFilm Lite eFilm Workstation with Modules is a software application that is used for viewing medical images. eFilm Workstation... on . FDA classified it as Class II and gives the reason as "A product issues happens for RF projection images. If user measures on RF image, .cal (Calibration) is not displayed and measurement is not correct. The .cal label is not displayed on the projection images after...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0707-2017
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Merge Healthcare Incorporated
Product
Merge, eFilm Workstation and eFilm Lite eFilm Workstation with Modules is a software application that is used for viewing medical images. eFilm Workstation with Modules receives digital images and data from various sources (including but not limited to CT, MR, US, RF units, computed and direct radiographic devices, secondary capture devices, scanners, imaging gateways or imaging sources). eFilm Workstation with Modules can be used to communicate, process and display medical images. Users have access to various image processing and measurement tools to assist them in viewing images. In addition, users can overlay templates on medical images to aid in preoperative planning. eFilm Workstation with Modules can be integrated with an institution s existing HIS or RIS for a fully integrated electronic patient record. Typical users of eFilm Workstation with Modules are trained medical professionals, including but not limited to radiologists, technologists and clinicians
Product code
LLZ System, Image Processing, Radiological
Reason for recall
A product issues happens for RF projection images. If user measures on RF image, .cal (Calibration) is not displayed and measurement is not correct. The .cal label is not displayed on the projection images after calibration is performed which should be present if the measurement is presented based on Pixel Spacing.
Root cause tag
software
FDA root cause
Software design
Action
Merge Healthcare sent an Urgent Medical Device Recall letter dated October 12, 2016, to all affected consignees. The letter described the Product, Issue, Potential Harm, Containment by the Customer / User, Actions by Merge, and Actions by Customers. Advised consignees to discontinue using the product, and identify & notify all customers. Requested consignees for a response to the notification. Customers with questions were instructed to send an email to [email protected].
Quantity in commerce
2163
Distribution
Worldwide Distribution - US: Nationwide including DC, GU, PR and Internationally to Canada, Argentina, Armenia, Australia, Austria, Belgium, Brazil, Chile, China, Colombia, Congo, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Ecuador, Estonia, Finland, France, Germany, Greece, Guatemala, hong Kong, Hungary, Iceland, India, Indonesia, Ireland, Israel, Italy, Japan, Kazakhstan, Latvia, Lebanon, Lithuania, Luxembourg, Malaysia, Mauritania, Mexico, Netherlands, New Caledonia, New Zealand, Norway, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Romania, Russia, Saudi Arabia, Serbia and Montenegro, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Vietnam, and Yemen.
Product codes and lots
EFilm Workstation and eFilm Lite versions 2.1, 2.1.2, 3.0, 3.1, 3.3.5, 3.4, 4.0, 4.0.1, 4.0.2, 4.0.3 and 4.1.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0707-2017
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0707-2017Class IITerminatedVoluntary: Firm initiated