Z-0348-2017Class II

Heater Cooler Unit Usage: The Heater-Cooler Unit (HCU 30) supplies temperature-controlled water for...

Maquet Cardiovascular Us Sales, LLC / initiated 2016-09-27 / Terminated / root cause: design

Maquet Cardiovascular Us Sales, Llc began a recall of Heater Cooler Unit Usage: The Heater-Cooler Unit (HCU 30) supplies temperature-controlled water for cardioplegia heat exchangers and for blood heat exchangers... on . FDA classified it as Class II and gives the reason as "There is the potential for contaminated water to enter parts of the Heater-Cooler Unit circuits. There is a potential risk of bacterial contamination including Mycobacteria species". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0348-2017
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Maquet Cardiovascular Us Sales, LLC
Product
Heater Cooler Unit Usage: The Heater-Cooler Unit (HCU 30) supplies temperature-controlled water for cardioplegia heat exchangers and for blood heat exchangers in extra corporeal circulation and for blankets with which patients can be warmed or cooled. The HCU 30 is intended to circulate water through heat exchange circuits to warm or cool a patient during short duration cardiopulmonary bypass procedures lasting 6 hours or less.
Product code
DWC Controller, Temperature, Cardiopulmonary Bypass
Reason for recall
There is the potential for contaminated water to enter parts of the Heater-Cooler Unit circuits. There is a potential risk of bacterial contamination including Mycobacteria species.
Root cause tag
design
FDA root cause
Device Design
Action
Maquet issued an Urgent Medical Device Recall letter/Return Response form dated September 27, 2016 to affected customers. The letter identified the affected product, problem and actions to be taken. Customer were instructed to complete the return response form via fax to 1-973-629-1518. For questions contact Maquet Customer Service at 1-888-627-8383 option 2 followed by option 2 Monday through Friday between 8 amd and 6 pm. EST.
Quantity in commerce
305 units
Distribution
Worldwide Distribution - US Nationwide
Product codes and lots
HCU 30 Base Unit 100-120 V, Model Number 704629 I 70103.4371 HCU 30 Base Unit Canadian/US, Model Number 705501 I 70103.4653 HCU 30 Base Unit 200-240 V, Model Number 939001 I 70102.8718 All serial numbers are affected.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0348-2017
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0348-2017Class IITerminatedVoluntary: Firm initiated