Z-0070-2017Class II

Medtronic Navigation O-arm O2 Surgical Imaging System Product Catalog Number: BI-700-02000

Medtronic Navigation, Inc / initiated 2016-09-22 / Terminated / root cause: software

Medtronic Navigation, Inc.-Littleton began a recall of Medtronic Navigation O-arm O2 Surgical Imaging System Product Catalog Number: BI-700-02000 Product Usage: The O-arm Surgical Imaging System is a... on . FDA classified it as Class II and gives the reason as "O-arm O2 Surgical Imaging System Spatial calibration may be erroneous in Stealth Station navigated images". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0070-2017
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Medtronic Navigation, Inc
Product
Medtronic Navigation O-arm O2 Surgical Imaging System Product Catalog Number: BI-700-02000 Product Usage: The O-arm Surgical Imaging System is a multi-dimensional surgical imaging platform that is designed for use in spine, orthopaedic, and trauma-related surgeries. It provides real-time, intra-operative imaging of a patients anatomy with high quality images and a large field-of-view in both two and three dimensions
Product code
OWB Interventional Fluoroscopic X-Ray System
Reason for recall
O-arm O2 Surgical Imaging System Spatial calibration may be erroneous in Stealth Station navigated images
Root cause tag
software
FDA root cause
Software design
Action
Medtronic Navigation contacted customers via telephone beginning September 23, 2016. Customers were informed of the affected product, problem and actions to be taken. Medtronic requests that users immediately discontinue the use of the O-Arm O2 Surgical Imaging system for automatic registration in Stealth Station navigated surgical procedures until further notice. Medtronic Service personnel have been scheduled to begin the necessary inspections and re-calibration activities on customer systems beginning on September 23, 2016.
Quantity in commerce
14 units
Distribution
Worldwide Distribution - US Nationwide in the states of CO, KA, NE, MI, MO, OH, WA and country of Switzerland
Product codes and lots
Software revisions 4.0.0, 4.0.1 and 4.0.2. System Serial Numbers: C0930  C0981  C0986  C1113 C1069 C1050 C1208 C1043 C1144 C1163 C1145 C1162 C1002 C1074

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0070-2017
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0070-2017Class IITerminatedVoluntary: Firm initiated