Medtronic Navigation O-arm O2 Surgical Imaging System Product Catalog Number: BI-700-02000
Medtronic Navigation, Inc.-Littleton began a recall of Medtronic Navigation O-arm O2 Surgical Imaging System Product Catalog Number: BI-700-02000 Product Usage: The O-arm Surgical Imaging System is a... on . FDA classified it as Class II and gives the reason as "O-arm O2 Surgical Imaging System Spatial calibration may be erroneous in Stealth Station navigated images". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0070-2017
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Medtronic Navigation, Inc
- Product
- Medtronic Navigation O-arm O2 Surgical Imaging System Product Catalog Number: BI-700-02000 Product Usage: The O-arm Surgical Imaging System is a multi-dimensional surgical imaging platform that is designed for use in spine, orthopaedic, and trauma-related surgeries. It provides real-time, intra-operative imaging of a patients anatomy with high quality images and a large field-of-view in both two and three dimensions
- Product code
- OWB Interventional Fluoroscopic X-Ray System
- Reason for recall
- O-arm O2 Surgical Imaging System Spatial calibration may be erroneous in Stealth Station navigated images
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Medtronic Navigation contacted customers via telephone beginning September 23, 2016. Customers were informed of the affected product, problem and actions to be taken. Medtronic requests that users immediately discontinue the use of the O-Arm O2 Surgical Imaging system for automatic registration in Stealth Station navigated surgical procedures until further notice. Medtronic Service personnel have been scheduled to begin the necessary inspections and re-calibration activities on customer systems beginning on September 23, 2016.
- Quantity in commerce
- 14 units
- Distribution
- Worldwide Distribution - US Nationwide in the states of CO, KA, NE, MI, MO, OH, WA and country of Switzerland
- Product codes and lots
- Software revisions 4.0.0, 4.0.1 and 4.0.2. System Serial Numbers: C0930 C0981 C0986 C1113 C1069 C1050 C1208 C1043 C1144 C1163 C1145 C1162 C1002 C1074
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0070-2017 | Class II | Terminated | Voluntary: Firm initiated |