510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0350-2017Class II

Tandem (Tm) Cocr Shell Uhmwpe Liner Bipolar, Ref Numbers: 71322040, 71322041, 71322042, 71322043, 71322044...

Smith & Nephew, Inc. / initiated 2016-08-29 / Terminated / root cause: manufacturing process

Smith & Nephew, Inc. began a recall of TANDEM (TM) COCR SHELL UHMWPE LINER BIPOLAR, REF NUMBERS: 71322040, 71322041, 71322042, 71322043, 71322044, 71322045, 71322046, 71322047, 71322048, 71322049... on . FDA classified it as Class II and gives the reason as "Some Bipolar shells were manufactured with an out of specification retainer groove". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0350-2017
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Smith & Nephew, Inc.
Product
Tandem (Tm) Cocr Shell Uhmwpe Liner Bipolar, Ref Numbers: 71322040, 71322041, 71322042, 71322043, 71322044, 71322045, 71322046, 71322047, 71322048, 71322049, 71322050, 71322051, 71322052, 71322053, 71322054, 71322055, 71322057
Product code
KWY Prosthesis, Hip, Hemi-, Femoral, Metal/Polymer, Cemented Or Uncemented
Reason for recall
Some Bipolar shells were manufactured with an out of specification retainer groove.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
The firm initiated their recall by FedEx and Email on 08/29/2016 to hospitals and distributors and requested the return of any in stock product. International consignees were notified by Email on 08/29/2016. Additionally, the firm notified the US physician of the issue by overnight letter on 09/08/2016.
Quantity in commerce
2352 units
Distribution
Nationwide and Australia, Belgium, Canada, China, Cyprus, France, Germany, India, Italy, Japan, Korea, Mexico, Netherlands, Poland, Portugal, Puerto Rico, Singapore, South Africa, Spain, Switzerland, Thailand, Turkey, United Arab Emirates
Product codes and lots
REF NUMBER 71322040 Serial Numbers: 14AM13511, 14AM13511A  REF NUMBER 71322041 Serial Numbers: 14AM21654  REF NUMBER 71322042 Serial Numbers: 13LM06353, 13MM17938, 13MM17940, 14AM01852, 14AM04660, 14AM04662, 13LM06354, 13MM17939   REF NUMBER 71322043 Serial Numbers: 13KM21801, 13LM09191, 13MM09504, 14AM01946, 14AM04663, 14BM15318, 13LM06356, 13LM09204, 13MM09507, 14AM02029, 14AM21664, 14BM15318A, 13LM06356A, 13MM02843, 14AM01855, 14AM02046, 14BM01137, 14BM19345  REF NUMBER 71322044 Serial Numbers:  13KM21825, 13LM09225, 13MM09510, 14AM02112, 14AM04670, 14BM08940A, 13KM24083, 13LM09236, 13MM09517, 14AM02118, 14AM13546, 13KM24083A, 13LM09236A, 13MM09517A, 14AM02120, 14BM01133, 13KM24084, 13MM02853, 13MM09517B, 14AM02120A, 14BM01133A, 13LM06357, 13MM02857, 13MM14125, 14AM04667, 14BM08938, 13LM06361, 13MM02857A, 13MM14135, 14AM04668, 14BM08940   REF NUMBER 71322045 Serial Numbers:  13KM17382, 13KM24087, 13MM02862A, 13MM17942, 14AM04672, 14BM08950, 13KM21826, 13LM00819, 13MM09520, 13MM17943, 14AM13549, 14BM08958,  13KM22753, 13LM09243, 13MM09520A, 13MM17943A, 14AM21666, 14BM15339, 13KM22755, 13LM09247, 13MM09523, 14AM02117, 14AM21669, 13KM24086,  13MM02862, 13MM17941, 14AM02117A, 14BM06955    REF NUMBER 71322046 Serial Numbers:  13KM17387, 13LM09274, 13LM11684, 13MM09524A, 14AM04682, 14BM15355, 13KM22756, 13LM09275, 13MM02888A, 13MM09526, 14AM13552, 14BM15358, 13KM22756A, 13LM09275A, 13MM02892, 13MM09527, 14AM21677, 14BM16787,  13KM24089, 13LM09278, 13MM02893, 13MM09991, 14AM21681, 14BM19253, 13LM00818, 13LM09281, 13MM02893A, 14AM02121, 14BM06952, 13LM00818A, 13LM11683, 13MM02897, 14AM02121A, 14BM12173, 13LM09258, 13LM11683A, 13MM09524, 14AM02124, 14BM12174   REF NUMBER 71322047 Serial Numbers:  13LM00809, 13LM00817B, 13LM11695, 13MM02901, 14AM21693, 14BM16794A, 13LM00809A, 13LM09285A, 13LM11700, 13MM02901B, 14BM12175, 14BM16795, 13LM00810, 13LM09289, 13LM14993, 13MM09528, 14BM12176, 14BM16795A, 13LM00816, 13LM09289A, 13LM15004, 13MM09529, 14BM16791, 13LM00817A,  13LM11689, 13MM02900, 14AM02127, 14BM16794   REF NUMBER 71322048 Serial Numbers:  13LM01881, 13LM01886, 13LM11730, 13MM02905, 14AM13555, 14BM08977, 13LM01882, 13LM11707, 13LM11730A, 13MM09994, 14AM16949, 14BM12177, 13LM01883, 13LM11707A, 13LM11731, 13MM17944, 14AM16951, 13LM01883A, 13LM11709, 13LM14017, 13MM17947, 14BM08976   REF NUMBER 71322049 Serial Numbers:  13LM01887, 13LM01889, 13LM11734, 13MM02906, 13MM13869A, 13MM14122A, 13LM01888, 13LM01890, 13LM14022, 13MM13868, 13MM14119, 14BM12178, 13LM01888A, 13LM11732, 13LM14023, 13MM13869, 13MM14122, 14BM12178A    REF NUMBER 71322050 Serial Numbers:  13LM11735, 13LM12855, 13LM15017, 13MM13870, 13MM17994, 13LM12820, 13LM12957, 13LM15751, 13MM13871, 13MM17995, 13LM12852, 13LM12960, 13LM15751A, 13MM13871A, 14BM01122  REF NUMBER 71322051 Serial Numbers:  13LM12961, 13LM12962A, 13LM14032A, 13LM14035, 13MM13872, 14BM16796, 13LM12961A, 13LM14025, 13LM14033, 13LM14037, 13MM13873, 14BM16796A, 13LM12962, 13LM14032, 13LM14034A, 13LM16186, 13MM17997   REF NUMBER 71322052 Serial Numbers:  13LM16191, 14AM02691, 14AM02694, 14AM04689, 14BM12179A, 14BM12182A, 13LM16192, 14AM02691A, 14AM04683, 14AM21695, 14BM12179B, 14BM12182B,  13LM16193, 14AM02692, 14AM04687, 14AM21695A, 14BM12181, 13LM16193A, 14AM02693, 14AM04688, 14BM12179, 14BM12182   REF NUMBER 71322053 Serial Numbers:  13LM16195, 14AM02695, 14AM02697, 14AM16934, 14AM21696, 14BM12183, 13LM16197,  14AM02696, 14AM06639, 14AM16935, 14AM21696A, 14BM12184   REF NUMBER 71322054 Serial Numbers:  13KM22757, 13LM16204, 14AM02699, 14AM02701B, 14AM16936, 14BM12186, 13KM22758, 13LM16204A, 14AM02700, 14AM02702, 14AM16937, 13LM16203, 13MM03205, 14AM02701, 14AM02702A, 14AM16937A, 13LM16203A, 14AM02698,  14AM02701A, 14AM02703, 14BM12185   REF NUMBER 71322055 Serial Numbers:  13LM16205, 14AM02710A, 14AM16940, 14BM01110A, 14BM12187, 14AM02710, 14AM16938, 14BM01110, 14BM01110B   REF NUMBER 71322057 Serial Numbers:  14AM16597, 14AM16597A

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0350-2017
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0350-2017Class IITerminatedVoluntary: Firm initiated
Z-0351-2017Class IITerminatedVoluntary: Firm initiated