sigma spectrum VOLUMETRIC INFUSION PUMP with Master Drug Library, V6 Spectrum Infusion Pump, Product Code...
Baxter Healthcare Corp. began a recall of sigma spectrum VOLUMETRIC INFUSION PUMP with Master Drug Library, V6 Spectrum Infusion Pump, Product Code 35700BAX and 35700BAXR, Unique Device Identifier GTIN... on . FDA classified it as Class II and gives the reason as "A correction was issued for the SIGMA Spectrum Infusion Pump with Master Drug Library in order to replace speakers installed in pumps or included in spare part rear case assemblies that have an increased potential to...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0670-2017
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Baxter Healthcare Corporation
- Product
- sigma spectrum VOLUMETRIC INFUSION PUMP with Master Drug Library, V6 Spectrum Infusion Pump, Product Code 35700BAX and 35700BAXR, Unique Device Identifier GTIN 00085412091570, Rx ONLY -- Sigma, LLC Medina, NY 14103 --- NOTE: The speakers are a part of the rear case assembly located in the back of the infusion pump. The speaker is intended to generate audible tones during each keystroke, and during alarm conditions. However, some speakers have a variation in their internal wiring that increases the potential for the speaker to fail prematurely. Spectrum Infusion Pumps with non-functional speakers will continue to display visual alarm indications, and will otherwise continue to operate as intended.
- Product code
- FRN Pump, Infusion
- Reason for recall
- A correction was issued for the SIGMA Spectrum Infusion Pump with Master Drug Library in order to replace speakers installed in pumps or included in spare part rear case assemblies that have an increased potential to fail prematurely.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Baxter sent an Urgent Medical Device Correction letter dated September 28, 2016, to all affected customers. On 10/27/2016, Baxter Healthcare sent a f/u UMDC Letter (dated 10/27/2016) to customers. The f/u letter contains the device correction plan which states that Baxter will be replacing at no charge the rear case assembly on all affected Sigma Spectrum Infusion Pumps, as well as affected stand-alone rear case assemblies shipped to customers as spare parts. For questions, please contact your Baxter sales representative, or Baxter Healthcare Medina at 800-356-3454 (choose option 1) Monday through Friday, between 7:00 am and 7:00 pm ET.
- Quantity in commerce
- 5,976 units (US: 5972, Canada: 4)
- Distribution
- Nationwide Distribution, Puerto Rico and Canada
- Product codes and lots
- Contact CDRH for list of affected serial numbers (5976 units)
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0670-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0671-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0672-2017 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.
