Z-0670-2017Class II

sigma spectrum VOLUMETRIC INFUSION PUMP with Master Drug Library, V6 Spectrum Infusion Pump, Product Code...

Baxter Healthcare Corporation / initiated 2016-09-28 / Terminated / root cause: manufacturing process

Baxter Healthcare Corp. began a recall of sigma spectrum VOLUMETRIC INFUSION PUMP with Master Drug Library, V6 Spectrum Infusion Pump, Product Code 35700BAX and 35700BAXR, Unique Device Identifier GTIN... on . FDA classified it as Class II and gives the reason as "A correction was issued for the SIGMA Spectrum Infusion Pump with Master Drug Library in order to replace speakers installed in pumps or included in spare part rear case assemblies that have an increased potential to...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0670-2017
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Baxter Healthcare Corporation
Product
sigma spectrum VOLUMETRIC INFUSION PUMP with Master Drug Library, V6 Spectrum Infusion Pump, Product Code 35700BAX and 35700BAXR, Unique Device Identifier GTIN 00085412091570, Rx ONLY -- Sigma, LLC Medina, NY 14103 --- NOTE: The speakers are a part of the rear case assembly located in the back of the infusion pump. The speaker is intended to generate audible tones during each keystroke, and during alarm conditions. However, some speakers have a variation in their internal wiring that increases the potential for the speaker to fail prematurely. Spectrum Infusion Pumps with non-functional speakers will continue to display visual alarm indications, and will otherwise continue to operate as intended.
Product code
FRN Pump, Infusion
Reason for recall
A correction was issued for the SIGMA Spectrum Infusion Pump with Master Drug Library in order to replace speakers installed in pumps or included in spare part rear case assemblies that have an increased potential to fail prematurely.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Baxter sent an Urgent Medical Device Correction letter dated September 28, 2016, to all affected customers. On 10/27/2016, Baxter Healthcare sent a f/u UMDC Letter (dated 10/27/2016) to customers. The f/u letter contains the device correction plan which states that Baxter will be replacing at no charge the rear case assembly on all affected Sigma Spectrum Infusion Pumps, as well as affected stand-alone rear case assemblies shipped to customers as spare parts. For questions, please contact your Baxter sales representative, or Baxter Healthcare Medina at 800-356-3454 (choose option 1) Monday through Friday, between 7:00 am and 7:00 pm ET.
Quantity in commerce
5,976 units (US: 5972, Canada: 4)
Distribution
Nationwide Distribution, Puerto Rico and Canada
Product codes and lots
Contact CDRH for list of affected serial numbers (5976 units)

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0670-2017
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0670-2017Class IITerminatedVoluntary: Firm initiated
Z-0671-2017Class IITerminatedVoluntary: Firm initiated
Z-0672-2017Class IITerminatedVoluntary: Firm initiated