Z-0292-2017Class II

Philips IntelliVue MX40 Patient Monitor: IntelliVue MX40 Smart-hopping (2.4GHz) Product Number: 865351...

Philips North America LLC / initiated 2016-09-19 / Terminated / root cause: software

Philips Electronics North America Corporation began a recall of Philips IntelliVue MX40 Patient Monitor: IntelliVue MX40 Smart-hopping (2.4GHz) Product Number: 865351 Worldwide Except USA Exchange part: 453564262491... on . FDA classified it as Class II and gives the reason as "Philips IntelliVue MX40 Patient Wearable Monitor Configuration Setting Disables Generation and Delivery of ECG Alarms to PIIC iX". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0292-2017
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Philips North America LLC
Product
Philips IntelliVue MX40 Patient Monitor: IntelliVue MX40 Smart-hopping (2.4GHz) Product Number: 865351 Worldwide Except USA Exchange part: 453564262491 453564262511 453564615311 453564615331 453564262571 453564262591 Exchange part: 453564262531 453564262551 453564262571 453564262591 453564615351 453564615371 453564467761 (Korea) 453564467781 (Korea) 453564451791 (Japan, Taiwan, Indonesia) 453564451811 (Japan, Taiwan, Indonesia) 453564467821 (Mexico) 453564467841 (China) 453564467861 (China) Worldwide except for the USA SW Revisions: B.05.28, B.05.29, and B.05.32 Indicated for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients. Intended for monitoring and recording of, and to generate alarms for, multiple physiological parameters of adults and pediatrics in hospital environments and during transport inside hospitals
Product code
MHX Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)
Reason for recall
Philips IntelliVue MX40 Patient Wearable Monitor Configuration Setting Disables Generation and Delivery of ECG Alarms to PIIC iX
Root cause tag
software
FDA root cause
Software design
Action
The firm, Philips Healthcare, sent a "Field Safety Notice (FSN)" dated 9/14/2016 to its customers. The notice described the product, problem and actions to be taken. The customers were instructed to make sure that the MX40 Patient Wearable Monitor local configuration setting 'Alarms On' is configured to its factory default until your software is upgraded; and follow the instructions listed in the letter. A Philips Healthcare representative will contact customers with affected devices to arrange for the installation of the software upgrade. If you need any further information or support concerning this issue, please contact Philips Customer Support at 1-800-722-9377.
Quantity in commerce
1824 units
Distribution
Worldwide Distribution-US (Nationwide) and countries of: Aruba, Australia, Austria, Belgium, Bulgaria, Canada, China, Czech Republic, Denmark, Finland, France, Germany, Ghana, Greece, Hong Kong, Iceland, India, Iraq, Ireland, Italy, Japan, Korea, Republic of, Malaysia, Mexico, Netherlands, New Zealand, Norway, Philippines, Poland, Portugal, Singapore, Slovenia, South Africa, Spain Sweden, Switzerland, Thailand, Turkey, United Arab Emirates and United Kingdom.
Product codes and lots
SW Revisions: B.05.28, B.05.29, and B.05.32

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0292-2017
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0291-2017Class IITerminatedVoluntary: Firm initiated
Z-0292-2017Class IITerminatedVoluntary: Firm initiated
Z-0293-2017Class IITerminatedVoluntary: Firm initiated