OEC Brivo 715 Prime, OEC Brivo 785 Essential, OEC Brivo 865 Advance
GE OEC Medical Systems, Inc began a recall of OEC Brivo 715 Prime, OEC Brivo 785 Essential, OEC Brivo 865 Advance Product Usage: The OEC Brivo Mobile C-Arm X-Ray Products are designed to provide digital... on . FDA classified it as Class II and gives the reason as "GE Healthcare Surgery announces a voluntary field action for the OEC Brivo 715 Prime, OEC Brivo 685 Essential, and OEC Brivo 865 Advance because the circuit board that controls vertical motion of the C-arm may fail...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0071-2017
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- GE Healthcare
- Product
- OEC Brivo 715 Prime, OEC Brivo 785 Essential, OEC Brivo 865 Advance Product Usage: The OEC Brivo Mobile C-Arm X-Ray Products are designed to provide digital spot film imaging and fluoroscopic image guidance for all adult and pediatric populations for orientations between patient anatomy and surgical instruments. The product is used for general surgical applications and musculoskeletal procedures to visualize, for example, implant localizations or needle positions for aspirations, injections or biopsy. The OEC Brivo is not indicated for interventional use.
- Product code
- OXO Image-Intensified Fluoroscopic X-Ray System, Mobile
- Reason for recall
- GE Healthcare Surgery announces a voluntary field action for the OEC Brivo 715 Prime, OEC Brivo 685 Essential, and OEC Brivo 865 Advance because the circuit board that controls vertical motion of the C-arm may fail, resulting in the C-Arm moving up or down without command.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- GE Healthcare sent an Urgent Medical Device Correction letter dated September 23, 2016 to customers. The letter informed customers of the affected product, safety issue, and safety instructions for immediate mitigation of the issue. GE Healthcare will correct all affected products at no cost to customers. A GE Healthcare representative will the customers to arrange for the correction. Please contact GE Healthcare Service at 1-800-437-1171 or your local Service Representative.
- Quantity in commerce
- 148 total
- Distribution
- Worldwide Distribution - US Nationwide in the states of AL, AZ, CA, FL, KS, MS, NJ, NV, NY, PA, TX, UT, and WV. No Candaian or Va/govt/military consignees. There are 133 other foreign consignees.
- Product codes and lots
- GE Brivo OEC715: Serial Numbers B2S15158 to B2S15164; B2S15166 to B2S15171; B2S15173 to B2S15195; B2S15197; B2S15200; B2S15202; B2S16046; B2S16050; B2S16051 GE Brivo OEC785: Serial Numbers B3S15236; B3S15241; B3S15242; B3S15246; B3S15248; B3S15250; B3S15253 to B3S15257; B3S15259 to B3S15338; B3S15353; B3S15357 to B3S15359; B3S16152 GE Brivo OEC865: Serial Numbers B4S15063 to B4S15070; B4S15072; B4S15073; B4S15075; B4S16029
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0071-2017 | Class II | Terminated | Voluntary: Firm initiated |