510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0366-2017Class II

Professional Hospital Supply, Inc Custom Packaging containing the Devon Light Glove Item Number Description...

Medtronic, Inc. / initiated 2016-09-29 / Terminated / root cause: manufacturing process

Covidien LLC began a recall of Professional Hospital Supply, Inc Custom Packaging containing the Devon Light Glove Item Number Description 380789003 SURE SET SINGLE BASIN PACK 400383006... on . FDA classified it as Class II and gives the reason as "Added Instructions for Use: After application, inspect the Light Glove for barrier integrity". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0366-2017
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Medtronic, Inc.
Product
Professional Hospital Supply, Inc Custom Packaging containing the Devon Light Glove Item Number Description 380789003 SURE SET SINGLE BASIN PACK 400383006 MAJOR BASIN PACK 400383007 MAJOR BASIN PACK 618275004 BASIN PACK Product Usage: A disposable, sterile, single-use cover for compatible surgical light handles
Product code
LRO General Surgery Tray
Reason for recall
Added Instructions for Use: After application, inspect the Light Glove for barrier integrity
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Medtronic issued a Correction Notice on September 29, 2016. Customers were notified via Federal Express and the letter informed customers of the addition of the following statement to the Instructions for Use (IFU): After application, inspect the Light Glove for barrier integrity. Questions or concerns, contact your Medtronic representative or Customer Service at (800) 962-9888
Quantity in commerce
4950 units
Distribution
Worldwide Distribution - US Nationwide Foreign: Canada: Andorra Austria Belgium Czech Republic Denmark Finland France Germany Greece Ireland Israel Italy Latvia Luxembourg Monaco Netherlands Norway Poland Portugal Russian Federation Saudi Arabia Slovenia South Africa Spain Sweden Switzerland United Arab Emirates United Kingdom
Product codes and lots
All lot codes in the following:  Item Number Description 380789003 SURE SET SINGLE BASIN PACK 400383006 MAJOR BASIN PACK 400383007 MAJOR BASIN PACK 618275004 BASIN PACK

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0366-2017
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0364-2017Class IITerminatedVoluntary: Firm initiated
Z-0365-2017Class IITerminatedVoluntary: Firm initiated
Z-0366-2017Class IITerminatedVoluntary: Firm initiated