ellman A CYNOSURE COMPANY Minor Surgery Bipolar Cable, Contents: Five (5) Disposable Bipolar Cables,STERILE...

Ellman International, Inc. / initiated 2016-09-07 / Terminated / root cause: manufacturing process

Ellman International, Inc. began a recall of ellman A CYNOSURE COMPANY Minor Surgery Bipolar Cable, Contents: Five (5) Disposable Bipolar Cables,STERILE The Minor Surgery Bipolar Cable is an accessory... on . FDA classified it as Class II and gives the reason as "Mislabeling of the "expiration date". The expiration date on the outer box and the individual pouches do not match. The expiration date on the box is correct with a date of 2018-02". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0748-2017
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Ellman International, Inc.
Product
ellman A CYNOSURE COMPANY Minor Surgery Bipolar Cable, Contents: Five (5) Disposable Bipolar Cables,STERILE The Minor Surgery Bipolar Cable is an accessory that is used to connect the bipolar handpieces to the RF generator for electrosurgical procedures that require the bipolar waveform.
Product code
GEI Electrosurgical, Cutting & Coagulation & Accessories
Reason for recall
Mislabeling of the "expiration date". The expiration date on the outer box and the individual pouches do not match. The expiration date on the box is correct with a date of 2018-02.
Root cause tag
manufacturing process
FDA root cause
Employee error
Action
Cynosure, Inc. dba Ellman distributed Urgent Medical Device Recall Notices dated September 6, 2016, and response forms to their customers via UPS overnight. Cynosure has identified an error in the Expiration Date on the label of the Minor Surgery Bipolar Cable pouch. The Expiration Date on the individual pouches that are provided in the 5 pack boxes do not match the Expiration Date on the box. The Expiration Date on the box is correct with a date of 2018-02, but the label on the individual pouches has a date of 2021-02. Actions To Be Taken - Return any cables subject to recall. You will be contacted by a Cynosure Customer Service Representative who will provide you with a RMA number for any cables that are being returned. Complete and return the response form via email to [email protected]. Cynosure will replace affected products at no charge upon receipt of the returned cables. Customers with questions should call 978-256-4200 or 800-886-2966. For questions regarding this recall call 516-267-6502.
Quantity in commerce
370 boxes (US - 159 boxes, OUS - 211 boxes)
Distribution
Worldwide Distribution - US including CA, FL, MA, MD, MI, MN, MS, OH, PA, VA, TX & WV; **International**Australia, Bangladesh, Greece, Italy, Japan, Lebanon, Russian Federation, Singapore, Spain & Vietman.
Product codes and lots
Lot Numbers 45572 (Box), (Individual packages) 48543, 50382, 48543, 50704, 50858 Exp Date - 2018-02

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0748-2017
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0748-2017Class IITerminatedVoluntary: Firm initiated