Z-1520-2017Class II

Alaris System PC unit, model no. 8000 and 8015. The central programming, monitoring and power supply...

CareFusion 303, Inc / initiated 2016-11-01 / Terminated

CareFusion 303, Inc. began a recall of Alaris System PC unit, model no. 8000 and 8015. The central programming, monitoring and power supply component for the Alaris infusion pump System on . FDA classified it as Class II and gives the reason as "Reports where the Low Battery alarm and/or the Very Low Battery alarm are not being triggered before the battery is discharged and all infusion channels are stopped". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-1520-2017
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
CareFusion 303, Inc
Product
Alaris System PC unit, model no. 8000 and 8015. The central programming, monitoring and power supply component for the Alaris infusion pump System.
Product code
FRN Pump, Infusion
Reason for recall
Reports where the Low Battery alarm and/or the Very Low Battery alarm are not being triggered before the battery is discharged and all infusion channels are stopped.
FDA root cause
Use error
Action
A customer letter dated 11/1/2016 will be sent to all customers to inform them that BD has identified an issue with the Alaris System PC unit. The letter informs the customers that BD has received reports for the Alaris PC units where the Low Battery alarm and/or the Very Low Battery alarm are not being triggered before the battery is discharged and all infusion channels are stopped. The letter informs the customers of the issue and recommended steps to take. Customers with recall related questions are instructed to contact BD Support Center at (888) 562-6018, 7-4pm PT, M-F, [email protected]. Customers with adverse event reports are instructed to contact Customer Advocacy at (888) 812-3266, 24hrs/day, 7days/wk, [email protected]. Customers with technical questions regarding the Alaris System are instructed to contact Technical Support at (888) 812-3229, 6am-5pm PT, M-F, [email protected].
Quantity in commerce
613,800 total units (575,221 units in US)
Distribution
Worldwide distribution. US nationwide, Europe, Australia, New Zealand, South Africa, Greater Asia, Middle East, and Canada.
Product codes and lots
all units

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1520-2017
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1520-2017Class IITerminatedVoluntary: Firm initiated