Z-0387-2017Class II

Roche COBAS INTEGRA c111 Analyzer, Chemistry (Photometric, Discrete), for clinical use

Roche Diagnostics GmbH / initiated 2016-09-27 / Terminated / root cause: software

Roche Diagnostics Operations, Inc. began a recall of Roche COBAS INTEGRA c111 Analyzer, Chemistry (Photometric, Discrete), for clinical use Product Usage: The Roche COBAS INTEGRA c111 analyzer is an in-vitro... on . FDA classified it as Class II and gives the reason as "cobas c 111 analyzers (catalog numbers 04777433001 and 04528778001) with software versions up to and including 4.20 may encounter the following alarm: 7002: 108000572, A software error occurred. This alarm is generated...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0387-2017
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Roche Diagnostics GmbH
Product
Roche COBAS INTEGRA c111 Analyzer, Chemistry (Photometric, Discrete), for clinical use Product Usage: The Roche COBAS INTEGRA c111 analyzer is an in-vitro diagnostic analyzer capable of performing clinical chemistry, specific protein and electrolyte tests. Analytes are measured photometrically or turbidimetrically. The analyzer also has an optional ISE module for measuring sodium, potassium and chloride.
Product code
JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
Reason for recall
cobas c 111 analyzers (catalog numbers 04777433001 and 04528778001) with software versions up to and including 4.20 may encounter the following alarm: 7002: 108000572, A software error occurred. This alarm is generated due to a measurement timing error. Under very rare conditions, the instrument may process two tests in the same cuvette if the run restarts. If a used cuvette is used again result of the test(s) will be erroneous. These erroneous results may not be flagged. Falsely low or high patient results may lead to incorrect diagnostic measures and medical therapeutic decisions. The medical risk depends on the parameter.
Root cause tag
software
FDA root cause
Software design
Action
Roche sent an URGENT MEDICAL DEVICE CORRECTION letter dated September 27, 2016 to the affected customers via courier. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. Customers were asked to complete the attached fax form and fax it to 1-866-808-1159. For questions contact the Roche Support Network Customer Support Center at 1-800-428-2336.
Quantity in commerce
139
Distribution
US Nationwide Distribution
Product codes and lots
04777433001 Cobas c111 with ISE 04528778001 Cobas c111 without ISE  cobas c 111 analyzers (catalog numbers 04777433001 and 04528778001) with software versions up to and including 4.20

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0387-2017
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0387-2017Class IITerminatedVoluntary: Firm initiated