Z-0390-2017Class II

Accessory Adapter Part number 600525A0

Maquet Cardiovascular Us Sales, LLC / initiated 2016-10-05 / Terminated / root cause: manufacturing process

Maquet Medical Systems USA began a recall of Accessory Adapter Part number 600525A0 Product Usage: designed for the mounting of MAQUET accessories with the dovetail guide interface immediately before... on . FDA classified it as Class II and gives the reason as "Maquet Inc. is initiating a voluntary field action on the Accessory Adapter part number 600525A0 due to a complaint in which a head rest mounted to the accessory adapter slipped out of the mountain rail". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0390-2017
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Maquet Cardiovascular Us Sales, LLC
Product
Accessory Adapter Part number 600525A0 Product Usage: designed for the mounting of MAQUET accessories with the dovetail guide interface immediately before, during and after surgical interventions as well as for examination and treatment.
Product code
FQO Table, Operating-Room, Ac-Powered
Reason for recall
Maquet Inc. is initiating a voluntary field action on the Accessory Adapter part number 600525A0 due to a complaint in which a head rest mounted to the accessory adapter slipped out of the mountain rail.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Maquet Medical Systems USA sent an Urgent Product Recall Medical Device Removal letter dated October 5, 2016 to the affected customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to: 1. Read this Medical Device Removal Notice and disseminate the information provided to all the users of the device. 2. Examine your inventory immediately to determine if you have the Accessory Adapter(s). If so, remove the affected products, quarantine them and keep them in a secure location for shipment back to Maquet. 3. Please fill out and sign the enclosed Medical Device Removal Response Form on page 4 and return it either via email to [email protected] or fax to 1-973-909-9927. For question contact your local MAQUET Representative or call MAQUET Customer Service at 1-888-627-8383 (Press option 2, then followed by option 3) Monday through Friday, between the hours of 8:00 a.m. and 5:00 p.m. EDT.
Quantity in commerce
9 Units (US) 141 Units (OUS)
Distribution
Worldwide Distribution - US Nationwide in the states of AZ, CA, DC, MI, MO, and PA
Product codes and lots
600525A0

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0390-2017
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0390-2017Class IITerminatedVoluntary: Firm initiated