Accessory Adapter Part number 600525A0
Maquet Medical Systems USA began a recall of Accessory Adapter Part number 600525A0 Product Usage: designed for the mounting of MAQUET accessories with the dovetail guide interface immediately before... on . FDA classified it as Class II and gives the reason as "Maquet Inc. is initiating a voluntary field action on the Accessory Adapter part number 600525A0 due to a complaint in which a head rest mounted to the accessory adapter slipped out of the mountain rail". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0390-2017
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Maquet Cardiovascular Us Sales, LLC
- Product
- Accessory Adapter Part number 600525A0 Product Usage: designed for the mounting of MAQUET accessories with the dovetail guide interface immediately before, during and after surgical interventions as well as for examination and treatment.
- Product code
- FQO Table, Operating-Room, Ac-Powered
- Reason for recall
- Maquet Inc. is initiating a voluntary field action on the Accessory Adapter part number 600525A0 due to a complaint in which a head rest mounted to the accessory adapter slipped out of the mountain rail.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- Maquet Medical Systems USA sent an Urgent Product Recall Medical Device Removal letter dated October 5, 2016 to the affected customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to: 1. Read this Medical Device Removal Notice and disseminate the information provided to all the users of the device. 2. Examine your inventory immediately to determine if you have the Accessory Adapter(s). If so, remove the affected products, quarantine them and keep them in a secure location for shipment back to Maquet. 3. Please fill out and sign the enclosed Medical Device Removal Response Form on page 4 and return it either via email to [email protected] or fax to 1-973-909-9927. For question contact your local MAQUET Representative or call MAQUET Customer Service at 1-888-627-8383 (Press option 2, then followed by option 3) Monday through Friday, between the hours of 8:00 a.m. and 5:00 p.m. EDT.
- Quantity in commerce
- 9 Units (US) 141 Units (OUS)
- Distribution
- Worldwide Distribution - US Nationwide in the states of AZ, CA, DC, MI, MO, and PA
- Product codes and lots
- 600525A0
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0390-2017 | Class II | Terminated | Voluntary: Firm initiated |