Z-0399-2017Class II
Merge LIS software
Merge Healthcare Incorporated / initiated 2015-03-05 / Terminated / root cause: software
Merge Healthcare, Inc. began a recall of Merge LIS software on . FDA classified it as Class II and gives the reason as "There is a potential for duplicate container numbers to be created for patients". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0399-2017
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Merge Healthcare Incorporated
- Product
- Merge LIS software.
- Product code
- JQP Calculator/Data Processing Module, For Clinical Use
- Reason for recall
- There is a potential for duplicate container numbers to be created for patients.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- The recalling firm issued a letter dated 1/27/2016 via e-mail on 1/30/2016. If no email address was available, the letters were issued via certified mail. Letters to nonresponders were issued 5/26/2016.
- Quantity in commerce
- 413 sites potentially have the affected software
- Distribution
- Distribution was made nationwide and to the Virgin Islands. Foreign distribution was made to the Bahamas.
- Product codes and lots
- Versions: 4.0.1 and earlier.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0399-2017 | Class II | Terminated | Voluntary: Firm initiated |