Z-0775-2017Class II

ANGIODYNAMICS Percutaneous Introducer Set under the following labels: 1) 5 F Stiffened Micro-Introducer Kit...

Boston Scientific Corporation / initiated 2016-09-28 / Terminated / root cause: design

Angiodynamics, Inc. began a recall of ANGIODYNAMICS Percutaneous Introducer Set under the following labels: 1) 5 F Stiffened Micro-Introducer Kit, 45CM NT/T ECHO B PG, Item Number H787065970515... on . FDA classified it as Class II and gives the reason as "Greatbatch Medical the manufacturer of the 4F and 5F sheath/dilator components has determined that the products listed in their September 7, 2016 Recall Notification have the potential for the rotating luer to detach...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0775-2017
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Boston Scientific Corporation
Product
ANGIODYNAMICS Percutaneous Introducer Set under the following labels: 1) 5 F Stiffened Micro-Introducer Kit, 45CM NT/T ECHO B PG, Item Number H787065970515, Catalog No.REF 06597051, 2) 5 F Stiffened Micro-Introducer Kit, 45CM NT/T ECHO S PG, Item Number H787065970535, Catalog No.REF 06597053, 3) 5 F Stiffened Micro-Introducer Kit, 45CM NT/T ECHO A PG, Item Number H787065970555, Catalog No.REF 06597055 Indicated for percutaneous introduction of guidewire or catheter into the vascular system following a small 21-gauge needle stick
Product code
DYB Introducer, Catheter
Reason for recall
Greatbatch Medical the manufacturer of the 4F and 5F sheath/dilator components has determined that the products listed in their September 7, 2016 Recall Notification have the potential for the rotating luer to detach from the dilator hub during use. AngioDynamics has confirmed that affected sheath/dilators (Greatbatch Model Numbers 10904-001 and 10904-002) have been included in packaged AngioDynamics Stiffened Micro-Introducer Kits.
Root cause tag
design
FDA root cause
Component design/selection
Action
Angiiodynamics sent an Urgent Voluntary Medical Device Recall letter dated September 28, 2016, to all affected customers. Recall notifications were delivered by Federal Express. Customers were instructed to segregate and return all affected devices to AngioDynamics and to complete and return the Reply Verification Tracking Form, provided in the recall notification by fax to 1-800-782-1357. Customers were encouraged to forward a copy of the recall notification to all sites to which have distributed affected product. If affected product is located in your institution, please call AngioDynamics Customer Service at 1-800-772-6446 between 8:00 a.m. and 7:00 p.m. (Monday Friday: Eastern Standard Time) to obtain a replacement or credit for your returned product. Promptly complete, sign and return the enclosed Reply Verification Tracking Form (even if you do not have any product to return); following the directions on this page and the Reply Verification Tracking Form via to [email protected]. Customers with questions should contact customer service at 1-800-772-6446.
Quantity in commerce
996 boxes
Distribution
Nationwide Distribution to AL, AR, CA, FL, GA, IL, KY, LA, MI, MN, MO, NC, OH, OK, SC, NY, ME, PA, MD, VA, WA, WI, WV, TN, and TX
Product codes and lots
Batch/Lot: 1) 5030537, 5037976, 5043444, 5059984, 5066413, 2) 5024545, 5031723, 5038659, 5042534, 5065538, 5071294, 5076428, 5082169, 3) 5022318, 5032456, 5034279, 5081986, 5082446

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0775-2017
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0772-2017Class IIITerminatedVoluntary: Firm initiated
Z-0773-2017Class IIITerminatedVoluntary: Firm initiated
Z-0774-2017Class IIITerminatedVoluntary: Firm initiated
Z-0775-2017Class IIITerminatedVoluntary: Firm initiated
Z-0776-2017Class IIITerminatedVoluntary: Firm initiated
Z-0777-2017Class IITerminatedVoluntary: Firm initiated
Z-0778-2017Class IIITerminatedVoluntary: Firm initiated