ANGIODYNAMICS 4F Stiffened Micro-Introducer, Item Number H787065971035, Catalog No.REF 06597103 Indicated for...
Angiodynamics, Inc. began a recall of ANGIODYNAMICS 4F Stiffened Micro-Introducer, Item Number H787065971035, Catalog No.REF 06597103 Indicated for percutaneous introduction of guidewire or... on . FDA classified it as Class II and gives the reason as "Greatbatch Medical the manufacturer of the 4F and 5F sheath/dilator components has determined that the products listed in their September 7, 2016 Recall Notification have the potential for the rotating luer to detach...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0778-2017
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Boston Scientific Corporation
- Product
- ANGIODYNAMICS 4F Stiffened Micro-Introducer, Item Number H787065971035, Catalog No.REF 06597103 Indicated for percutaneous introduction of guidewire or catheter into the vascular system following a small 21-gauge needle stick
- Product code
- DYB Introducer, Catheter
- Reason for recall
- Greatbatch Medical the manufacturer of the 4F and 5F sheath/dilator components has determined that the products listed in their September 7, 2016 Recall Notification have the potential for the rotating luer to detach from the dilator hub during use. AngioDynamics has confirmed that affected sheath/dilators (Greatbatch Model Numbers 10904-001 and 10904-002) have been included in packaged AngioDynamics Stiffened Micro-Introducer Kits.
- Root cause tag
- design
- FDA root cause
- Component design/selection
- Action
- Angiiodynamics sent an Urgent Voluntary Medical Device Recall letter dated September 28, 2016, to all affected customers. Recall notifications were delivered by Federal Express. Customers were instructed to segregate and return all affected devices to AngioDynamics and to complete and return the Reply Verification Tracking Form, provided in the recall notification by fax to 1-800-782-1357. Customers were encouraged to forward a copy of the recall notification to all sites to which have distributed affected product. If affected product is located in your institution, please call AngioDynamics Customer Service at 1-800-772-6446 between 8:00 a.m. and 7:00 p.m. (Monday Friday: Eastern Standard Time) to obtain a replacement or credit for your returned product. Promptly complete, sign and return the enclosed Reply Verification Tracking Form (even if you do not have any product to return); following the directions on this page and the Reply Verification Tracking Form via to [email protected]. Customers with questions should contact customer service at 1-800-772-6446.
- Quantity in commerce
- 4 boxes
- Distribution
- Nationwide Distribution to AL, AR, CA, FL, GA, IL, KY, LA, MI, MN, MO, NC, OH, OK, SC, NY, ME, PA, MD, VA, WA, WI, WV, TN, and TX
- Product codes and lots
- Batch/Lot: 1) 5034283
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0772-2017 | Class III | Terminated | Voluntary: Firm initiated | ||
| Z-0773-2017 | Class III | Terminated | Voluntary: Firm initiated | ||
| Z-0774-2017 | Class III | Terminated | Voluntary: Firm initiated | ||
| Z-0775-2017 | Class III | Terminated | Voluntary: Firm initiated | ||
| Z-0776-2017 | Class III | Terminated | Voluntary: Firm initiated | ||
| Z-0777-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0778-2017 | Class III | Terminated | Voluntary: Firm initiated |