Z-0700-2017Class II

Model 37751 Recharger

Medtronic Neuromodulation / initiated 2016-10-03 / Terminated / root cause: software

Medtronic Neuromodulation began a recall of Model 37751 Recharger Product Usage: The Medtronic 37751 Recharger is a hand-held device used to charge the battery in a patient s neurostimulator. It includes... on . FDA classified it as Class II and gives the reason as "Medtronic has identified an increased number of complaints from customers involving reports of Rechargers (Medtronic Model 37751 Recharger, which is included in Models 37754 and 97754 Charging Systems,Spinal Cord...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0700-2017
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Medtronic Neuromodulation
Product
Model 37751 Recharger Product Usage: The Medtronic 37751 Recharger is a hand-held device used to charge the battery in a patient s neurostimulator. It includes a display to provide information on the charging system. The charging system consists of the Model 37751 Recharger, an AC power supply and power cord (Model 37761) and an antenna.
Product code
MHY Stimulator, Electrical, Implanted, For Parkinsonian Tremor
Reason for recall
Medtronic has identified an increased number of complaints from customers involving reports of Rechargers (Medtronic Model 37751 Recharger, which is included in Models 37754 and 97754 Charging Systems,Spinal Cord Stimulation)that are in an unresponsive error state, where the Recharger is non-functional with a blank display screen and is beeping every 5 seconds. Medtronic has determined all Rechargers manufactured starting in November 2014 (indicated by serial numbers beginning with NKA4 or NKU4) are more susceptible to this error state.
Root cause tag
software
FDA root cause
Software change control
Action
Consignees were sent a series of documents which included an introductory Medtronic and "Urgent Medical Device Correction" letters dated September 2016. The urgent letters included one for Deep Brain Stimulation (DBS) and one for Spinal Cord Stimulation(SCS) . Each letter was addressed to Healthcare Professional. Only physicians associated with both SCS and DBS therapy received the cover letter. The letters provided information on "Background, Issues Mitigation, Recommendation, Additional Information and Patient Notification (DBS). Requested consignees to complete and return the Physician /HCP Reply Form. For questions contact Medtronic Technical Services at 1-800-707-0933 weekdays 7 am to 6pm, CT.
Quantity in commerce
42,887
Distribution
Worldwide Distribution - US Nationwide including DC OUS: Canada and others.
Product codes and lots
serial numbers beginning with NKA4 or NKU4

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0700-2017
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0700-2017Class IITerminatedVoluntary: Firm initiated