Z-0950-2017Class I
Alaris Pump Module, Model No. 8100 and AIL sensor kits, P/N 147083-102 and P/N 49000221. The Alaris Pump...
CareFusion 303, Inc / initiated 2016-12-02 / Terminated
CareFusion 303, Inc. began a recall of Alaris Pump Module, Model No. 8100 and AIL sensor kits, P/N 147083-102 and P/N 49000221. The Alaris Pump module is a large volume infusion pump offered under... on . FDA classified it as Class I and gives the reason as "During an infusion, a false Air-in-Line (AIL) alarm would cause the infusion to be interrupted". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0950-2017
- Classification
- Class I
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- CareFusion 303, Inc
- Product
- Alaris Pump Module, Model No. 8100 and AIL sensor kits, P/N 147083-102 and P/N 49000221. The Alaris Pump module is a large volume infusion pump offered under the Alaris System. The Alaris Pump module will deliver medication and fluids using IV administration sets for continuous or intermittent delivery through clinically acceptable routes of administration such as intravenous or epidural
- Product code
- FRN Pump, Infusion
- Reason for recall
- During an infusion, a false Air-in-Line (AIL) alarm would cause the infusion to be interrupted.
- FDA root cause
- Component change control
- Action
- Carefusion (BD) sent a Medical Device Safety Notification letter dated December 2, 2016, to all affected customers. The letter informed customers that BD has identified an issue with the Alaris Pump module. The letter informs the customers that the firm has received reports of Air-In-Line (AIL) alarms that have occurred when no air is observed in the line. Customers were informed of the issue and of the potential risk involved. The letter informs the customers of the actions for users and additional actions for biomedical engineering. Customers with questions regarding recall related questions were instructed to contact BD Support Center at (888) 562-6018, 7-4pm PT, M-F or email [email protected]. Customers with adverse event reports are instructed to contact Customer Advocacy at (888) 812-3266, 24 hrs/day, 7days/wk. or email [email protected]. For questions regarding this recall call 858-617-2000.
- Quantity in commerce
- 382,635 units
- Distribution
- Nationwide Distribution
- Product codes and lots
- All products manufactured or serviced in the time period October 2011 through June 2015.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0950-2017 | Class I | Terminated | Voluntary: Firm initiated |