Z-0315-2017Class II

da Vinci Xi" Surgical System, model number IS4000, A70_P5x with P5 Software; General and Plastic Surgery: The...

Intuitive Surgical, Inc. / initiated 2016-10-05 / Terminated / root cause: software

Intuitive Surgical, Inc. began a recall of da Vinci Xi" Surgical System, model number IS4000, A70_P5x with P5 Software; General and Plastic Surgery: The Intuitive Surgical Endoscope Instrument Control... on . FDA classified it as Class II and gives the reason as "Intuitive Surgical has identified a software anomaly in the da Vinci Xi P5 software that can result in unexpected master movement and potential instrument tip movement under certain circumstances". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0315-2017
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Intuitive Surgical, Inc.
Product
da Vinci Xi" Surgical System, model number IS4000, A70_P5x with P5 Software; General and Plastic Surgery: The Intuitive Surgical Endoscope Instrument Control System (da Vinci Surgical System, Model IS4000) is intended to assist in the accurate control of Intuitive Surgical Endoscopic Instruments including rigid endoscopes, blunt and sharp endoscopic dissectors, scissors, scalpels, forceps/pick-ups, needle holders, endoscopic retractors, electrocautery and accessories for endoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximations, ligation, electrocautery, suturing and delivery and placement of microware and cryogenic ablation probes and accessories, during urologic surgical procedures, general laparoscopic surgical procedures, gynecologic laparoscopic surgical procedures, general thoracoscopic surgical procedures and thoracoscopically-assisted cardiotomy procedures.
Product code
NAY System, Surgical, Computer Controlled Instrument
Reason for recall
Intuitive Surgical has identified a software anomaly in the da Vinci Xi P5 software that can result in unexpected master movement and potential instrument tip movement under certain circumstances
Root cause tag
software
FDA root cause
Software design
Action
Urgent Medical Device Correction letters were sent on October 12, 2016 by traceable method to all affected customers. Letters notify customers of the issue and provided instructions to prevent it from happening until ISI software can install new software upgrade in 1st Q 2017.
Quantity in commerce
677 devices
Distribution
Australia, Austria, Belgium, Colombia, Cyprus, Czech Republic, Denmark, Finland, France, Germany, Hong Kong, India, Ireland, Israel, Italy, Japan, Monaco, Netherlands, Norway, Portugal, Puerto Rico, Qatar, Romania, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey, United Kingdom, United States
Product codes and lots
All systems manufactured with or upgraded to P5 software.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0315-2017
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0315-2017Class IITerminatedVoluntary: Firm initiated