da Vinci Xi" Surgical System, model number IS4000, A70_P5x with P5 Software; General and Plastic Surgery: The...
Intuitive Surgical, Inc. began a recall of da Vinci Xi" Surgical System, model number IS4000, A70_P5x with P5 Software; General and Plastic Surgery: The Intuitive Surgical Endoscope Instrument Control... on . FDA classified it as Class II and gives the reason as "Intuitive Surgical has identified a software anomaly in the da Vinci Xi P5 software that can result in unexpected master movement and potential instrument tip movement under certain circumstances". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0315-2017
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Intuitive Surgical, Inc.
- Product
- da Vinci Xi" Surgical System, model number IS4000, A70_P5x with P5 Software; General and Plastic Surgery: The Intuitive Surgical Endoscope Instrument Control System (da Vinci Surgical System, Model IS4000) is intended to assist in the accurate control of Intuitive Surgical Endoscopic Instruments including rigid endoscopes, blunt and sharp endoscopic dissectors, scissors, scalpels, forceps/pick-ups, needle holders, endoscopic retractors, electrocautery and accessories for endoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximations, ligation, electrocautery, suturing and delivery and placement of microware and cryogenic ablation probes and accessories, during urologic surgical procedures, general laparoscopic surgical procedures, gynecologic laparoscopic surgical procedures, general thoracoscopic surgical procedures and thoracoscopically-assisted cardiotomy procedures.
- Product code
- NAY System, Surgical, Computer Controlled Instrument
- Reason for recall
- Intuitive Surgical has identified a software anomaly in the da Vinci Xi P5 software that can result in unexpected master movement and potential instrument tip movement under certain circumstances
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Urgent Medical Device Correction letters were sent on October 12, 2016 by traceable method to all affected customers. Letters notify customers of the issue and provided instructions to prevent it from happening until ISI software can install new software upgrade in 1st Q 2017.
- Quantity in commerce
- 677 devices
- Distribution
- Australia, Austria, Belgium, Colombia, Cyprus, Czech Republic, Denmark, Finland, France, Germany, Hong Kong, India, Ireland, Israel, Italy, Japan, Monaco, Netherlands, Norway, Portugal, Puerto Rico, Qatar, Romania, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey, United Kingdom, United States
- Product codes and lots
- All systems manufactured with or upgraded to P5 software.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0315-2017 | Class II | Terminated | Voluntary: Firm initiated |