Xhibit Central Station, Model 96102. Intended use is to provide clinicians with central monitoring of adult...
Spacelabs Healthcare Inc began a recall of Xhibit Central Station, Model 96102. Intended use is to provide clinicians with central monitoring of adult, pediatric and neonatal patient data of patients... on . FDA classified it as Class II and gives the reason as "The firm received reports of telemetry SpO2 numerics dropping off the Xhibit Central display. Desaturation, high, and low limit alarms work normally". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0332-2017
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- GE Healthcare
- Product
- Xhibit Central Station, Model 96102. Intended use is to provide clinicians with central monitoring of adult, pediatric and neonatal patient data of patients connected to networked Spacelabs Healthcare patient monitors and telemetry transmitters.
- Product code
- MHX Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)
- Reason for recall
- The firm received reports of telemetry SpO2 numerics dropping off the Xhibit Central display. Desaturation, high, and low limit alarms work normally.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- The firm sent the Urgent - Medical Device Correction letter, dated 18 October 2016, via priority service, return receipt requested, to all U.S. customers on 10/18/16. The firm will e-mail a customer letter (translated as necessary) to all international subsidiaries and distributors of record on October 25, 2016. Xhibit Central Stations used without Xhibit Telemetry Receivers, are not affected. Customers were asked to circulate the letter to all person affected and place a copy with the appropriate monitor user manual until corrections have been implemented. Spacelabs Healthcare will contact affected consignees to schedule a convenient time for Spacelabs to update the affected Xhibit Central Station(s) with a new software version 1.1.6 at no cost. Customers with questions can call Technical Support at 1-800-522-7025 and select 2.
- Quantity in commerce
- 348 units (292 units in the US and 56 units outside the US)
- Distribution
- Worldwide Distribution -- USA, to AL, CA, CT, FL, GA, IA, ID, IL, IN, KY, MI, MN, MS, NC, NE, NJ, NY, OH, OK, PA, RI, SC, SD, TX, VA, WA, WY, and Puerto Rico; and, to the countries of AUSTRALIA, CANADA, CHILE, CZECH REPUBLIC, FRANCE, MEXICO, NETHERLANDS, PANAMA, SAUDI ARABIA, SWITZERLAND, TAIWAN, and UNITED KINGDOM.
- Product codes and lots
- Software version 1.1.5, Affected Serial Numbers in the US ************************************* ADDITIONAL FIVE UNITS ADDED: SLA0216F01796; SLA0216F01808; SLA1215F01638; SLA1215F01645; SLA1215F01652; SLA0114F00445, SLA0114F00446, SLA0114F00447, SLA0115F00917, SLA0115F00918, SLA0115F00919, SLA0115F00921, SLA0115F00923, SLA0115F00924, SLA0115F00927, SLA0115F00928, SLA0115F00934, SLA0115F00955, SLA0115F00961, SLA0115F00968, SLA0115F00969, SLA0115F00972, SLA0115F00976, SLA0116F01718, SLA0116F01720, SLA0116F01722, SLA0214F00489, SLA0216F01807, SLA0216F01809, SLA0216F01813, SLA0216F01821, SLA0216F01824, SLA0216F01837, SLA0216F01840, SLA0216F01844, SLA0314F00534, SLA0314F00535, SLA0314F00540, SLA0314F00543, SLA0314F00547, SLA0314F00548, SLA0314F00549, SLA0314F00552, SLA0314F00561, SLA0314F00567, SLA0414F00568, SLA0414F00585, SLA0416F01850, SLA0416F01853, SLA0416F01857, SLA0416F01860, SLA0416F01875, SLA0416F01876, SLA0416F01883, SLA0416F01885, SLA0416F01889, SLA0416F01895, SLA0416F01923, SLA0516F01969, SLA0516F01970, SLA0516F01972, SLA0516F01974, SLA0516F01975, SLA0516F01976, SLA0516F01977, SLA0516F01978, SLA0516F01981, SLA0516F01982, SLA0516F01985, SLA0516F01986, SLA0516F01987, SLA0516F01992, SLA0516F02002, SLA0516F02005, SLA0614F00631, SLA0614F00654, SLA0615F01188, SLA0615F01194, SLA0615F01197, SLA0615F01198, SLA0615F01199, SLA0615F01213, SLA0615F01237, SLA0615F01238, SLA0615F01239, SLA0615F01240, SLA0615F01244, SLA0615F01248, SLA0615F01261, SLA0714F00678, SLA0714F00688, SLA0714F00694, SLA0714F00702, SLA0714F00703, SLA0715F01274, SLA0715F01310, SLA0715F01314, SLA0715F01325, SLA0715F01331, SLA0715F01334, SLA0715F01346, SLA0715F01347, SLA0716F02009, SLA0716F02011, SLA0716F02013, SLA0716F02015, SLA0716F02016, SLA0716F02017, SLA0716F02020, SLA0716F02021, SLA0716F02022, SLA0716F02028, SLA0716F02034, SLA0716F02042, SLA0716F02043, SLA0716F02044, SLA0716F02045, SLA0814F00713, SLA0815F01371, SLA0815F01372, SLA0815F01373, SLA0815F01374, SLA0815F01375, SLA0815F01376, SLA0815F01379, SLA0815F01380, SLA0815F01381, SLA0815F01382, SLA0815F01383, SLA0815F01384, SLA0815F01386, SLA0815F01387, SLA0815F01388, SLA0815F01390, SLA0815F01391, SLA0815F01395, SLA0815F01397, SLA0815F01400, SLA0815F01402, SLA0815F01403, SLA0815F01404, SLA0815F01405, SLA0815F01406, SLA0815F01407, SLA0815F01408, SLA0815F01409, SLA0815F01412, SLA0815F01415, SLA0815F01418, SLA0815F01419, SLA0815F01424, SLA0816F02048, SLA0816F02050, SLA0816F02051, SLA0816F02052, SLA0816F02053, SLA0816F02054, SLA0816F02055, SLA0816F02056, SLA0816F02057, SLA0816F02058, SLA0816F02059, SLA0816F02060, SLA0816F02061, SLA0816F02062, SLA0816F02063, SLA0816F02065, SLA0816F02066, SLA0816F02067, SLA0816F02074, SLA0816F02075, SLA0816F02076, SLA0816F02077, SLA0816F02078, SLA0816F02081, SLA0816F02082, SLA0816F02083, SLA0816F02085, SLA0816F02087, SLA0816F02089, SLA0816F02091, SLA0816F02092, SLA0816F02093, SLA0816F02094, SLA0816F02095, SLA0816F02096, SLA0816F02097, SLA0816F02098, SLA0816F02099, SLA0816F02100, SLA0816F02101, SLA0816F02102, SLA0816F02103, SLA0816F02104, SLA0816F02105, SLA0816F02106, SLA0816F02107, SLA0915F01439, SLA0915F01442, SLA0915F01447, SLA0915F01451, SLA0915F01452, SLA0915F01457, SLA0915F01459, SLA0915F01462, SLA0915F01468, SLA0915F01474, SLA0915F01493, SLA0915F01496, SLA0915F01497, SLA0915F01498, SLA0915F01499, SLA0915F01501, SLA0915F01503, SLA0915F01506, SLA0915F01507, SLA0915F01508, SLA0915F01510, SLA0915F01511, SLA0915F01512, SLA0915F01513, SLA0915F01514, SLA0916F02108, SLA0916F02110, SLA0916F02111, SLA0916F02112, SLA0916F02114, SLA0916F02115, SLA0916F02117, SLA0916F02121, SLA0916F02122, SLA0916F02123, SLA0916F02126, SLA0916F02127, SLA0916F02130, SLA0916F02131, SLA0916F02132, SLA0916F02133, SLA0916F02134, SLA0916F02136, SLA0916F02137, SLA0916F02138, SLA0916F02139, SLA0916F02141, SLA0916F02142, SLA0916F02143, SLA0916F02145, SLA0916F02146, SLA0916F02147, SLA0916F02148, SLA0916F02149, SLA0916F02150, SLA0916F02153, SLA0916F02154, SLA0916F02155, SLA0916F02156, SLA0916F02158, SLA0916F02159, SLA0916F02160, SLA0916F02162, SLA0916F02163, SLA0916F02165, SLA0916F02166, SLA1015F01525, SLA1015F01528, SLA1015F01530, SLA1015F01540, SLA1015F01551, SLA1015F01560, SLA1114F00795, SLA1114F00820, SLA1115F01578, SLA1115F01579, SLA1115F01582, SLA1115F01584, SLA1115F01593, SLA1115F01594, SLA1115F01595, SLA1115F01596, SLA1115F01597, SLA1115F01599, SLA1115F01602, SLA1115F01603, SLA1115F01608, SLA1115F01609, SLA1214F00858, SLA1214F00871, SLA1214F00876, SLA1214F00881, SLA1214F00885, SLA1214F00886, and SLA1215F01632. ***********Affected Serial Numbers OUTSIDE the US ************************** SLA0114F00430, SLA0114F00437, SLA0114F00438, SLA0114F00439, SLA0114F00440, SLA0114F00443, SLA0115F00975, SLA0116F01724, SLA0314F00528, SLA0314F00541, SLA0414F00596, SLA0416F01872, SLA0416F01877, SLA0416F01880, SLA0612F00235, SLA0715F01270, SLA0715F01271, SLA0715F01273, SLA0715F01283, SLA0715F01335, SLA0716F02023, SLA0716F02025, SLA0716F02029, SLA0716F02031, SLA0716F02035, SLA0716F02036, SLA0716F02037, SLA0716F02040, SLA0716F02041, SLA0716F02047, SLA0815F01431, SLA0816F02049, SLA0816F02064, SLA0816F02069, SLA0816F02070, SLA0816F02071, SLA0816F02072, SLA0816F02073, SLA0816F02079, SLA0816F02080, SLA0816F02084, SLA0816F02086, SLA0916F02109, SLA0916F02118, SLA0916F02119, SLA0916F02120, SLA0916F02128, SLA0916F02129, SLA0916F02140, SLA0916F02161, SLA1014F00776, SLA1113F00348, SLA1113F00368, SLA1115F01604, SLA1115F01605, and SLA1213F00389.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0332-2017 | Class II | Terminated | Voluntary: Firm initiated |