Z-0788-2017Class I

Medtronic SynchroMed II Implantable Drug Infusion System. This system includes: Model 8870 Software...

Medtronic Neuromodulation / initiated 2016-10-03 / Terminated / root cause: labeling and UDI

Medtronic Neuromodulation began a recall of Medtronic SynchroMed II Implantable Drug Infusion System. This system includes: Model 8870 Software Application Card, Model 8840 N'Vision Clinician Programmer... on . FDA classified it as Class I and gives the reason as "Medtronic is following up to a May 2013 communication regarding the Priming Bolus function for the SynchroMed Infusion System. Medtronic is updating the Model 8870 software application card (to version AAU01) and the...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-0788-2017
Classification
Class I
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Medtronic Neuromodulation
Product
Medtronic SynchroMed II Implantable Drug Infusion System. This system includes: Model 8870 Software Application Card, Model 8840 N'Vision Clinician Programmer, and Model 8637 SynchroMed II Implantable Drug Infusion Pump (supplied in 20 ml or 40 ml reservoir size.) Product Usage: The SynchroMed II Programmable Pump is indicated when patient therapy requires the chronic infusion of the following drugs or fluids: Chronic intraspinal (epidural and intrathecal) infusion of preservative-free morphine sulfate sterile solution in the treatment of chronic intractable pain, Chronic intrathecal infusion of preservative-free ziconotide sterile solution for the management of severe chronic pain, Chronic intrathecal infusion of Lioresal Intrathecal (baclofen Injection ) in the management of serve spasticity, Chronic Intravascular Infusion of Floxuridine (FUDR) or methotrexate for the treatment of primary or metastatic cancer
Product code
LKK Pump, Infusion, Implanted, Programmable
Reason for recall
Medtronic is following up to a May 2013 communication regarding the Priming Bolus function for the SynchroMed Infusion System. Medtronic is updating the Model 8870 software application card (to version AAU01) and the SynchroMed pump labeling to address the priming bolus issue.
Root cause tag
labeling and UDI
FDA root cause
Under Investigation by firm
Action
Medtronic Representatives began visiting consignees on 10/3/2016 to replace previous version Model 8870 software application cards, contained within the model 8840 N Vision Clinician Programmer, with the updated version (new version is AAU01). Beginning October 20, 2016, Implanting and Following physicians were sent an Urgent Medical Device Correction letter dated September 2016, . This letter to physicians is informing them of the updated Model 8870 software application card, and changes to SynchroMed infusion system labeling. A response form is requested to be returned to Medtronic. Customers with questions were instructed to contact Medtronic Neuromodulation Technical Services at 1-800-707-0933.
Quantity in commerce
22,298 software cards
Distribution
Worldwide Distribution - (US Nationwide) Internationally to Germany, Greece, Australia, United Kingdom, Switzerland, Italy, Norway, France, Canada, and Malta.
Product codes and lots
software application cards with versions other than AAU01

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0788-2017
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0788-2017Class ITerminatedVoluntary: Firm initiated