Z-0589-2017Class II

Advia Chemistry Calibrator For in vitro diagnostic use in the calibration of ADVIA Chemistry systems for the...

Siemens Healthcare Diagnostics Inc. / initiated 2016-10-07 / Terminated / root cause: manufacturing process

Siemens Healthcare Diagnostics, Inc. began a recall of Advia Chemistry Calibrator For in vitro diagnostic use in the calibration of ADVIA Chemistry systems for the following methods: Albumin (ALB, ALB_c), Direct... on . FDA classified it as Class II and gives the reason as "Siemens has confirmed that the DBIL_2 Assigned Calibrator Values (ACV) currently assigned to the affected lots may cause, on average, a positive proportional bias of approximately 41% on patient samples compared to the...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0589-2017
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
Advia Chemistry Calibrator For in vitro diagnostic use in the calibration of ADVIA Chemistry systems for the following methods: Albumin (ALB, ALB_c), Direct Bilirubin (DBIL_2), Total Bilirubin (TBIL_2), Calcium (CA, CA_c, CA_2, CA_2c), Cholesterol (CHOL), Creatinine (CREA, CREA_2, CRE_2c, ECRE_2), Glucose Hexakinase (GLUH), Glucose Oxidase (GLUO), Iron (IRON_2), Lithium (LITH), Magnesium (MG, MG_c), Phosphorus - inorganic (IP, IP_c), Total Protein (TP, TP_c), Triglycerides (TRIG), Urea Nitrogen (UN, UN_c), Uric Acid (UA, UA_c)
Product code
JJY Multi-Analyte Controls, All Kinds (Assayed)
Reason for recall
Siemens has confirmed that the DBIL_2 Assigned Calibrator Values (ACV) currently assigned to the affected lots may cause, on average, a positive proportional bias of approximately 41% on patient samples compared to the reference method for Direct Bilirubin measurement.
Root cause tag
manufacturing process
FDA root cause
Under Investigation by firm
Action
Siemens issued an Urgent Medical Device Correction in the US and an Urgent Field Safety Notice Outside the US beginning on October 6, 2016. Customers were notified of the issue and instructed to use the reassigned DBIL_2 calibrator value for Chemistry Calibrator Lot Numbers 680726, 680726A, 680726B, 680726C and 680726D. Customers were also requested to complete and return the Effectiveness Check questionnaire attached to the Urgent Medical Device Correction letter within 30 days by faxing to the Customer Care Center at 312-275-7795. For questions regarding this recall call 914-631-8000.
Quantity in commerce
10,614 units
Distribution
Worldwide Distribution - US (Nationwide) Internationally to Canada, Mexico, Algeria, Australia, Austria, Bangladesh, Belgium, Brazil, Chile, Cyprus, Denmark, Egypt, Ecuador, Finland, France, Germany, Great Britain, Hong Kong, Hungary, Italy, Japan, Lithuania, Malaysia, Mexico, Netherlands, New Zealand, Norway, Pakistan, Paraguay, Peru, Poland, Portugal, Russian Fed. Slovakia, Slovenia, Spain, Sweden, Switzerland, Thailand, Turkey, Vatican City and Vietnam
Product codes and lots
Device Listing # D033650 Model# SMN# 10312279 Lot Numbers: 680726, 680726A, 680726B, 680726C, 680726D

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0589-2017
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0589-2017Class IITerminatedVoluntary: Firm initiated