Z-0648-2017Class II

AQUIOS Lyse Reagent Kit, Catalog No. B23538

Siemens Healthcare Diagnostics Inc. / initiated 2016-10-19 / Terminated

Beckman Coulter Inc. began a recall of AQUIOS Lyse Reagent Kit, Catalog No. B23538 Product Usage: AQUIOS Lysing Reagent Kit is used as part of the AQUIOS flow cytometer system. The kit consists of... on . FDA classified it as Class II and gives the reason as "Beckman Coulter is recalling the AQUIOS Lyse Reagent Kit because it may lead to an increased generation of notifications and flags when used on the AQUIOS CL when running the AQUIOS Tetra application. The issue could...". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0648-2017
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
AQUIOS Lyse Reagent Kit, Catalog No. B23538 Product Usage: AQUIOS Lysing Reagent Kit is used as part of the AQUIOS flow cytometer system. The kit consists of two reagents used by AQUIOS flow cytometers to prepare whole blood samples for analysis of white blood cells.
Product code
OYE Flow Cytometric Reagents And Accessories.
Reason for recall
Beckman Coulter is recalling the AQUIOS Lyse Reagent Kit because it may lead to an increased generation of notifications and flags when used on the AQUIOS CL when running the AQUIOS Tetra application. The issue could lead to an increase in the number of samples requiring review and re-runs with a potential delay to the reporting of results. Depending on the specimen age, a new blood specimen may be required.
FDA root cause
Under Investigation by firm
Action
Beckman Coulter sent an Urgent Medical Device Recall letter dated October 19, 2016 to customers. The letter identified the affected product, problem and actions to be taken. Customers are instructed to complete and return the enclosed Response Form within 10 days. Customers with any questions regarding the notice are instructed to contact our Customer Support Center http://www.beckmancoulter.com/customersupport/support; 1-800-369-0333 in the US and Canada; outside the US and Canada contact local Beckman Coulter Representative.
Quantity in commerce
2367 units total (207 units in US)
Distribution
Worldwide Distribution - US Nationwide and the countries of Belgium, Botswana, Canada, Chile, Columbia, Czech Republic, Denmark, Finland, France, Germany, Ghana, Italy, Malaysia, Netherlands, Poland, Romania, Slovakia, South Africa, Spain, Sweden, Switzerland, United Kingdom and Zambia
Product codes and lots
6040014K 6040015K

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0648-2017
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0648-2017Class IITerminatedVoluntary: Firm initiated