Perkin Elmer RoboRack 25ul, Clear, Non-Conductive, Filter Tips, Pre-sterilized Item Number: 6000689. Supplied...
PerkinElmer, Inc began a recall of Perkin Elmer RoboRack 25ul, Clear, Non-Conductive, Filter Tips, Pre-sterilized Item Number: 6000689. Supplied for use with JANUS Automated Workstations... on . FDA classified it as Class II and gives the reason as "Filter Tips molded incorrectly and may not seal to the Varispan tip adaptor properly causing inaccurate aspiration and dispensing". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0747-2017
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Revvity Health Sciences, Inc.
- Product
- Perkin Elmer RoboRack 25ul, Clear, Non-Conductive, Filter Tips, Pre-sterilized Item Number: 6000689. Supplied for use with JANUS Automated Workstations: Product Number: AJS4001 AJM4001 AJI4001, AJL4001, AJS8001 , AJM8001, AJI8001, AJL8001, AJM4G01 , AJI4G01 , AJL4G01, AJM8G01, AJI8G01, AJL8G01, AJM4M01, AJI4M01 , AJL4M01, AJM8M01, AJI8M01, AJL8M01, AGS4NGS The JANUS Automated Workstation is an automated, programmable liquid handling instrument intended for use in clinical settings to facilitate the pipetting and dilution of solutions in sample preparation work flows. The NGS Express is an automated, programmable liquid handling instrument suitable for use in clinical settings to facilitate the pipetting and dilution of solutions in sample preparation workflows
- Product code
- JQW Station, Pipetting And Diluting, For Clinical Use
- Reason for recall
- Filter Tips molded incorrectly and may not seal to the Varispan tip adaptor properly causing inaccurate aspiration and dispensing
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- PerkinElmer initiated recall on June 24, 2016 to the Product Manager and was delivered to the customer by email facilitated through PerkinElmer Sales Team.Customers were requested to take the following actions until PerkinElmer's corrective measures are completed: Inspect customer inventory of the 6000689 25 Clear Filter Pipette Tip and identify if the affected lots are present. Complete and return the attached Response Form with the required information. Destroy all quantities of tips from the affected lots. Upon receipt of the return Response Form, a shipment of replacement tips will be shipped.
- Quantity in commerce
- 595 trays (571,200 tips)
- Distribution
- Worldwide Distribution - US including CA, MA, NY, IA, OR , PA, WA and Internationally to Belgium, Canada, Finland, France, Germany, Japan, Netherlands, Singapore, and UK
- Product codes and lots
- Lot Numbers:15470394, 15471775, 1601067, 16071147
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0747-2017 | Class II | Terminated | Voluntary: Firm initiated |