Z-0376-2017Class II

Philips Healthcare System X-ray Tomography Computed

Philips North America LLC / initiated 2016-03-15 / Terminated

Philips Electronics North America Corporation began a recall of Philips Healthcare System X-ray Tomography Computed on . FDA classified it as Class II and gives the reason as "It was discovered the patient support tabletop may fail to move in the horizontal direction when it is commanded to do so during CT scans due to a potential mechanical linkage failure". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0376-2017
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Philips North America LLC
Product
Philips Healthcare System X-ray Tomography Computed :
Product code
JAK System, X-Ray, Tomography, Computed
Reason for recall
It was discovered the patient support tabletop may fail to move in the horizontal direction when it is commanded to do so during CT scans due to a potential mechanical linkage failure.
FDA root cause
Radiation Control for Health and Safety Act
Action
Philips planned action: 1. Philips issued a customer notification letter on October 13, 2016. The letter includes a statement that the corrections will be made free of charge. 2. Philips will initiate a field corrective action to address the issue. 3. Philips will contact the customer and schedule an inspection to determine if the unit requires repair. 4. If repair is required, the couch horizontal linkage assembly will be replaced. For further questions please call (978) 659-3000.
Quantity in commerce
1363
Distribution
USA(nationwide) Distribution to the states of : AZ, NY, GA, UT, TN, TX, AR, NV, Il, FL, MS, LA, MD, MO, MA, OK, KY, WY, SC, ND, NE, DE, MI, WA, CT, RI, KS, OR, IN, PA, CA, NJ, NC, MN, AL and IA.
Product codes and lots
Ingenuity CT, Ingenuity Core, Ingenuity Core 128.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0376-2017
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0375-2017Class IITerminatedFDA Mandated
Z-0376-2017Class IITerminatedFDA Mandated