Z-0655-2017Class II
SURESOUND22US Multipack (US Distribution) containing 2 packs of six devices. SURESOUND22 Multipack (OUS...
Hologic, Inc. / initiated 2016-10-28 / Terminated / root cause: packaging
Hologic, Inc began a recall of SURESOUND22US Multipack (US Distribution) containing 2 packs of six devices. SURESOUND22 Multipack (OUS Distribution) containing 2 packs of six devices. For... on . FDA classified it as Class II and gives the reason as "Breach in sterile barrier compromising product sterility". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.
Recall record
- Recall number
- Z-0655-2017
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Hologic, Inc.
- Product
- SURESOUND22US Multipack (US Distribution) containing 2 packs of six devices. SURESOUND22 Multipack (OUS Distribution) containing 2 packs of six devices. For uterine sound
- Product code
- HHM Sound, Uterine
- Reason for recall
- Breach in sterile barrier compromising product sterility
- Root cause tag
- packaging
- FDA root cause
- Package design/selection
- Action
- Hologic notified customers in the United States through Certified letter on 10/28/16. The letter contains instructions with the option of returning or destroying the product, provides contact information, and includes a postage paid return card for the consignee to complete and return to Hologic.Questions or require clarification of the response process please contact 1- 800-442-9892 and selection option 6 and then option 2 for surgical product technical support. For questions regarding this recall call 508-263-8920.
- Quantity in commerce
- 600 units (US) ; 24 units (OUS)
- Distribution
- Worldwide Distribution - US (Nationwide) and Internationally to Finland and Austria.
- Product codes and lots
- Lot Code US: 16E25RC 16E25RD 16F16R Lot Code OUS: 16E25RB