Z-0665-2017Class II

Merge Hemo software. Merge Hemo monitors, measures, and records physiologic data from a human patient...

Merge Healthcare Incorporated / initiated 2014-04-30 / Terminated / root cause: software

Merge Healthcare, Inc. began a recall of Merge Hemo software. Merge Hemo monitors, measures, and records physiologic data from a human patient undergoing a cardiac catheterization procedure. The... on . FDA classified it as Class II and gives the reason as "The application may crash during the cath lab procedure". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0665-2017
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Merge Healthcare Incorporated
Product
Merge Hemo software. Merge Hemo monitors, measures, and records physiologic data from a human patient undergoing a cardiac catheterization procedure. The Monitoring System is for the monitoring of vital parameters including ECG, SpO2, invasive blood pressure, temperature, NIBP and CO2, and for the evaluation of resting ECG, arrhythmias, ST-segments and cardiac output. Some systems are built and designed to measure End Tidal CO2. The system is intended for use in hospital cardiac catheterization laboratories
Product code
DQK Computer, Diagnostic, Programmable
Reason for recall
The application may crash during the cath lab procedure.
Root cause tag
software
FDA root cause
Software design
Action
Merge sent an Important Alert letter to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers with questions were instructed to contact Merge Support at 866-261-5761 or [email protected]. For questions regarding this recall call 877-741-5369.
Quantity in commerce
192 sites potentially have the affected software
Distribution
Nationwide Distribution
Product codes and lots
Versions 9.10, 9.20.0, 9.20.2, 9.30.0, 9.40.0, 9.40.1, and 9.40.2 with Windows 7 Operating System

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0665-2017
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0665-2017Class IITerminatedVoluntary: Firm initiated