510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0682-2017Class II

STCKERT HeaterCooler System 3T; Item Number 160281, HeaterCooler 3T, 240V/60Hz; Item Number 160282...

LivaNova USA, Inc. / initiated 2014-08-01 / Terminated / root cause: manufacturing process

Sorin Group USA, Inc. began a recall of STCKERT HeaterCooler System 3T; Item Number 160281, HeaterCooler 3T, 240V/60Hz; Item Number 160282, HeaterCooler 3T, 208V/60Hz; Item Number 160285... on . FDA classified it as Class II and gives the reason as "On August 1, 2014 Sorin Group USA, Inc. informed consignees via letter to review the disinfection and monitoring practices and to strictly adhere to the instructions in the Instructions for Use (IFU)". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0682-2017
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
LivaNova USA, Inc.
Product
STCKERT HeaterCooler System 3T; Item Number 160281, HeaterCooler 3T, 240V/60Hz; Item Number 160282, HeaterCooler 3T, 208V/60Hz; Item Number 160285, HeaterCooler 3T, 120V/ 60Hz The Heater Cooler 3T is used with a Stckert S3 heart lung machine and / or any other heart lung machine featuring a separate temperature control for extracorporeal perfusion of durations up to 6 hours
Product code
DWC Controller, Temperature, Cardiopulmonary Bypass
Reason for recall
On August 1, 2014 Sorin Group USA, Inc. informed consignees via letter to review the disinfection and monitoring practices and to strictly adhere to the instructions in the Instructions for Use (IFU).
Root cause tag
manufacturing process
FDA root cause
Material/Component Contamination
Action
Sorin Group sent an Important Information letter dated August 1, 2014, to all affected consignees. The letter identified the product, the problem, and the action to be taken by then consignee. Consignees were instructed to review the disinfection and monitoring practices and to strictly adhere to the instructions in the Instructions for Use (IFU). Consignees with questions were instructed to call 303-467-6527 or 303-467-6503. For questions regarding this recall call 281-228-7260.
Quantity in commerce
1,618 units in the US; 129 units outside of US
Distribution
Worldwide Distribution - US (Nationwide and Puerto Rico) and Internationally to Canada, Colombia, Ecuador, El Salvador, Germany, Honduras, Mexico, Saudi Arabia, Trinidad, and Tobago.
Product codes and lots
All Serial Numbers

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0682-2017
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0682-2017Class IITerminatedVoluntary: Firm initiated