Z-0768-2017Class II

MicroScan LabPro Data Management System. Intended to manage both microbial identification (ID) and...

Siemens Healthcare Diagnostics Inc. / initiated 2016-10-25 / Terminated / root cause: design

Beckman Coulter Inc. began a recall of MicroScan LabPro Data Management System. Intended to manage both microbial identification (ID) and antimicrobial agent susceptibility testing (AST) data... on . FDA classified it as Class II and gives the reason as "Beckman Coulter has received and confirmed reports of an intermittent and unexpected behavior when loading new panels into a WalkAway instrument using LabPro Data Management System version 4.42. The issue could cause...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0768-2017
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
MicroScan LabPro Data Management System. Intended to manage both microbial identification (ID) and antimicrobial agent susceptibility testing (AST) data generated from MicroScan instruments or manually entered microbiology test results, for use by trained laboratory personnel.
Product code
OUG Medical Device Data System
Reason for recall
Beckman Coulter has received and confirmed reports of an intermittent and unexpected behavior when loading new panels into a WalkAway instrument using LabPro Data Management System version 4.42. The issue could cause workflow interruption with a potential of delaying reporting results due to the inability to begin processing new panels.
Root cause tag
design
FDA root cause
Device Design
Action
An Urgent Medical Device Recall letter dated 10/25/16 was sent to all their customers to inform them that Beckman Coulter has received and confirmed reports of an intermittent and unexpected behavior when loading new panels into a WalkAway instrument using LabPro Data Management System version 4.42. Customers are informed that the issue could cause WalkAway workflow interruption with a potential of delaying reporting results due to the inability to begin processing new panels. The letter informs the customers of the actions to be taken and the resolution for the product on recall. Customers are instructed to complete and return the enclosed Response Form within 10 days. Customers with any questions regarding the notice are instructed to contact their Customer Support representative: http://www.beckmancoulter.com or call 1-800-677-7226, option 1 in the United States.
Quantity in commerce
131 units total (130 units in US)
Distribution
Worldwide Distribution -- United States and Mexico.
Product codes and lots
Software Version 4.42, Part No. 6000-0050, 6000-0051, 6000-0052, 6000-0053, 6000-0054, and 6000-0060.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0768-2017
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0768-2017Class IITerminatedVoluntary: Firm initiated