Z-0692-2017Class II

SOMATOM Force, System x-ray, tomography, computed

Siemens Medical Solutions USA, Inc. / initiated 2016-10-20 / Terminated / root cause: software

Siemens Medical Solutions USA, Inc began a recall of SOMATOM Force, System x-ray, tomography, computed on . FDA classified it as Class II and gives the reason as "Siemens is providing software update version VA50A_SP3 to address the software bugs that were identified through normal field monitoring and the Global Complaint Handling Process. Correction for the problems are as...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0692-2017
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
SOMATOM Force, System x-ray, tomography, computed
Product code
JAK System, X-Ray, Tomography, Computed
Reason for recall
Siemens is providing software update version VA50A_SP3 to address the software bugs that were identified through normal field monitoring and the Global Complaint Handling Process. Correction for the problems are as follows: 1. Correction to volumetric misrepresentations of high contrast objects when using ADMIRE. 2. Correction to highly sporadic scan aborts due to temporarily tube currents at 0mA. 3. Correction to missing Microsoft Hotfixes (MS16-001, MS15-135, MS15-088, MS15-048). 4. Correction to highly sporadic image artifacts caused by error signals from high voltage chain at extremely high tube currents. 5. Correction to highly sporadic incomplete image reconstruction due to temporary slow hard disk write speed. 6. Correction to highly sporadic scan aborts due to temporarily high disk load of system disk. 7. Correction to highly sporadic image artifacts generated by scatter correction in case of tube arcings.
Root cause tag
software
FDA root cause
Under Investigation by firm
Action
Siemens sent a Customer Information Letter dated November 2, 2016, to customers affected by the recall to inform them of a release of a software update that will provide bug-fixes to improve system performance. Customers will be contacted to arrange a date for onsite installation. For further questions, please call (610) 448-6471.
Quantity in commerce
48 units distributed in U.S.
Distribution
Nationwide Distribution.
Product codes and lots
Model Number 10742326, Serial Numbers: 75708,75709,75607,75617,75550,75526 75512,75567,75536,75711,75514,75600,75563,75510,75454,75540,75571,75500 75527,75475,75535,75481,75482,75710,75460,75476,75588,75493,75458,75528 75467,75602,75487,75464,75478,75604,75513,75532,75555,75524,75515,75559 75610,75501,75570,75616,75576,75605

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0692-2017
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0692-2017Class IITerminatedVoluntary: Firm initiated