Z-0703-2017Class II

LCS Complete Knee Revision System VVC Insert

Depuy Orthopaedics Inc. / initiated 2016-11-08 / Terminated / root cause: manufacturing process

DePuy Orthopaedics, Inc. began a recall of LCS Complete Knee Revision System VVC Insert Product Usage: LCS VVC Insert is used in Knee Revision Surgery on . FDA classified it as Class II and gives the reason as "DePuy Orthopaedics, Inc. is voluntarily recalling one lot of the LCS Complete Knee Revision System VVC Insert Small 22.5mm, due to the insert being manufactured outside of specifications, which prevents the insert from...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0703-2017
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Depuy Orthopaedics Inc.
Product
LCS Complete Knee Revision System VVC Insert Product Usage: LCS VVC Insert is used in Knee Revision Surgery
Product code
NJL Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Metal/Polymer, Mobile Bearing
Reason for recall
DePuy Orthopaedics, Inc. is voluntarily recalling one lot of the LCS Complete Knee Revision System VVC Insert Small 22.5mm, due to the insert being manufactured outside of specifications, which prevents the insert from properly seating in the mating tibial tray.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
DePuy Orthopaedics, Inc. initiated a voluntary recall on 11/8/2016 via letter mailed to the customer. Only one device was distributed in the U.S., and this was removed from the hospital on 21-Oct-2016. This recall does not affect any other lots of the LCS Complete Knee Revision System VVC Insert Small 22.5mm devices. The recall notice provides instructions for notifying the medical facility that received, used, or purchased the affected lot of the LCS Complete Knee Revision System VVC Insert Small 22.5mm. The purpose of this device recall is to notify medical professionals of the possible effects of using the affected device and to communicate that there are no recalled devices distributed at this time. The customer was asked to do the following: Complete the Reconciliation Form and return to your DePuy Orthopaedics Sales Consultant or fax to 574-371-4939 or email to [email protected] within five (5) business-days of this notice. " Retain a copy of the completed Reconciliation Form and this notice in your files. " Forward this notice to others in your facility that need to be informed. Effectiveness will be determined by the return of one US Medical Facility Reconciliation Form. For questions about device recall information provided, please contact Kim Earle, Senior Recall Coordinator, at 574-371-4917 or [email protected] or [email protected].
Quantity in commerce
1
Distribution
US Distribution in the state of TN
Product codes and lots
LCS COMP VVC INS SM 22.5MM, Product Code: 129424122, Lot# C08294, GTIN: (01)10603295024484, Expiry: 31-Mar-2021

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0703-2017
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0703-2017Class IITerminatedVoluntary: Firm initiated